Effect of tube current on computed tomography radiomic features.

Publication ,  Journal Article
Mackin, D; Ger, R; Dodge, C; Fave, X; Chi, P-C; Zhang, L; Yang, J; Bache, S; Dodge, C; Jones, AK; Court, L
Published in: Scientific reports
February 2018

Variability in the x-ray tube current used in computed tomography may affect quantitative features extracted from the images. To investigate these effects, we scanned the Credence Cartridge Radiomics phantom 12 times, varying the tube current from 25 to 300 mA∙s while keeping the other acquisition parameters constant. For each of the scans, we extracted 48 radiomic features from the categories of intensity histogram (n = 10), gray-level run length matrix (n = 11), gray-level co-occurrence matrix (n = 22), and neighborhood gray tone difference matrix (n = 5). To gauge the size of the tube current effects, we scaled the features by the coefficient of variation of the corresponding features extracted from images of non-small cell lung cancer tumors. Variations in the tube current had more effect on features extracted from homogeneous materials (acrylic, sycamore wood) than from materials with more tissue-like textures (cork, rubber particles). Thirty-eight of the 48 features extracted from acrylic were affected by current reductions compared with only 2 of the 48 features extracted from rubber particles. These results indicate that variable x-ray tube current is unlikely to have a large effect on radiomic features extracted from computed tomography images of textured objects such as tumors.

Duke Scholars

Published In

Scientific reports

DOI

EISSN

2045-2322

ISSN

2045-2322

Publication Date

February 2018

Volume

8

Issue

1

Start / End Page

2354

Related Subject Headings

  • Tomography, X-Ray Computed
  • Phantoms, Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Electricity
  • Carcinoma, Non-Small-Cell Lung
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Mackin, D., Ger, R., Dodge, C., Fave, X., Chi, P.-C., Zhang, L., … Court, L. (2018). Effect of tube current on computed tomography radiomic features. Scientific Reports, 8(1), 2354. https://doi.org/10.1038/s41598-018-20713-6
Mackin, Dennis, Rachel Ger, Cristina Dodge, Xenia Fave, Pai-Chun Chi, Lifei Zhang, Jinzhong Yang, et al. “Effect of tube current on computed tomography radiomic features.Scientific Reports 8, no. 1 (February 2018): 2354. https://doi.org/10.1038/s41598-018-20713-6.
Mackin D, Ger R, Dodge C, Fave X, Chi P-C, Zhang L, et al. Effect of tube current on computed tomography radiomic features. Scientific reports. 2018 Feb;8(1):2354.
Mackin, Dennis, et al. “Effect of tube current on computed tomography radiomic features.Scientific Reports, vol. 8, no. 1, Feb. 2018, p. 2354. Epmc, doi:10.1038/s41598-018-20713-6.
Mackin D, Ger R, Dodge C, Fave X, Chi P-C, Zhang L, Yang J, Bache S, Jones AK, Court L. Effect of tube current on computed tomography radiomic features. Scientific reports. 2018 Feb;8(1):2354.

Published In

Scientific reports

DOI

EISSN

2045-2322

ISSN

2045-2322

Publication Date

February 2018

Volume

8

Issue

1

Start / End Page

2354

Related Subject Headings

  • Tomography, X-Ray Computed
  • Phantoms, Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Electricity
  • Carcinoma, Non-Small-Cell Lung