Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification.

Publication ,  Journal Article
Vrbaski, S; Bache, S; Rajagopal, J; Samei, E
Published in: Med Phys
September 2023

BACKGROUND: Quantitative imaging techniques, such as virtual monochromatic imaging (VMI) and iodine quantification (IQ), have proven valuable diagnostic methods in several specific clinical tasks such as tumor and tissue differentiation. Recently, a new generation of computed tomography (CT) scanners equipped with photon-counting detectors (PCD) has reached clinical status. PURPOSE: This work aimed to investigate the performance of a new photon-counting CT (PC-CT) in low-dose quantitative imaging tasks, comparing it to an earlier generation CT scanner with an energy-integrating detector dual-energy CT (DE-CT). The accuracy and precision of the quantification across size, dose, material types (including low and high iodine concentrations), displacement from iso-center, and solvent (tissue background) composition were explored. METHODS: Quantitative analysis was performed on two clinical scanners, Siemens SOMATOM Force and NAEOTOM Alpha using a multi-energy phantom with plastic inserts mimicking different iodine concentrations and tissue types. The tube configurations in the dual-energy scanner were 80/150Sn kVp and 100/150Sn kVp, while for PC-CT both tube voltages were set to either 120 or 140 kVp with photon-counting energy thresholds set at 20/65 or 20/70 keV. The statistical significance of patient-related parameters in quantitative measurements was examined using ANOVA and pairwise comparison with the posthoc Tukey honest significance test. Scanner bias was assessed in both quantitative tasks for relevant patient-specific parameters. RESULTS: The accuracy of IQ and VMI in the PC-CT was comparable between standard and low radiation doses (p < 0.01). The patient size and tissue type significantly affect the accuracy of both quantitative imaging tasks in both scanners. The PC-CT scanner outperforms the DE-CT scanner in the IQ task in all cases. Iodine quantification bias in the PC-CT (-0.9 ± 0.15 mg/mL) at low doses in our study was comparable to that of DE-CT (range -2.6 to 1.5 mg/mL, published elsewhere) at a 1.7× higher dose, but the dose reduction severely biased DE-CT (4.72 ± 0.22 mg/mL). The accuracy in Hounsfield units (HU) estimation was comparable for 70 and 100 keV virtual imaging between scanners, but PC-CT was significantly underestimating virtual 40 keV HU values of dense materials in the phantom representing the extremely obese population. CONCLUSIONS: The statistical analysis of our measurements reveals better IQ at lower radiation doses using new PC-CT. Although VMI performance was mostly comparable between the scanners, the DE-CT scanner quantitatively outperformed PC-CT when estimating HU values in the specific case of very large phantoms and dense materials, benefiting from increased X-ray tube potentials.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

September 2023

Volume

50

Issue

9

Start / End Page

5421 / 5433

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Tomography Scanners, X-Ray Computed
  • Photons
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Iodine
  • Humans
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Vrbaski, S., Bache, S., Rajagopal, J., & Samei, E. (2023). Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification. Med Phys, 50(9), 5421–5433. https://doi.org/10.1002/mp.16583
Vrbaski, Stevan, Steve Bache, Jayasai Rajagopal, and Ehsan Samei. “Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification.Med Phys 50, no. 9 (September 2023): 5421–33. https://doi.org/10.1002/mp.16583.
Vrbaski S, Bache S, Rajagopal J, Samei E. Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification. Med Phys. 2023 Sep;50(9):5421–33.
Vrbaski, Stevan, et al. “Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification.Med Phys, vol. 50, no. 9, Sept. 2023, pp. 5421–33. Pubmed, doi:10.1002/mp.16583.
Vrbaski S, Bache S, Rajagopal J, Samei E. Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification. Med Phys. 2023 Sep;50(9):5421–5433.

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

September 2023

Volume

50

Issue

9

Start / End Page

5421 / 5433

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Tomography Scanners, X-Ray Computed
  • Photons
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Iodine
  • Humans
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis