Chronic in vivo testing of the Penn State infant ventricular assist device.

Publication ,  Journal Article
Weiss, WJ; Carney, EL; Clark, JB; Peterson, R; Cooper, TK; Nifong, TP; Siedlecki, CA; Hicks, D; Doxtater, B; Lukic, B; Yeager, E; Reibson, J ...
Published in: ASAIO J
2012

The Penn State Infant Ventricular Assist Device (VAD) is a 12-14 ml stroke volume pneumatically actuated pump, with custom Björk-Shiley monostrut valves, developed under the National Heart, Lung, and Blood Institute Pediatric Circulatory Support program. In this report, we describe the seven most recent chronic animal studies of the Infant VAD in the juvenile ovine model, with a mean body weight of 23.5 ± 4.1 kg. The goal of 4-6 weeks survival was achieved in five of seven studies, with support duration ranging from 5 to 41 days; mean 26.1 days. Anticoagulation was accomplished using unfractionated heparin, and study animals were divided into two protocol groups: the first based on a target activated partial thromboplastin time of 1.5-2 times normal, and a second group using a target thromboelastography R-time of two times normal. The second group required significantly less heparin, which was verified by barely detectable heparin activity (anti-Xa). In both groups, there was no evidence of thromboembolism except in one animal with a chronic infection and fever. Device thrombi were minimal and were further reduced by introduction of the custom valve. These results are consistent with results of adult VAD testing in animals and are encouraging given the extremely low levels of anticoagulation in the second group.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

ASAIO J

DOI

EISSN

1538-943X

Publication Date

2012

Volume

58

Issue

1

Start / End Page

65 / 72

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Sheep
  • Prosthesis Design
  • Partial Thromboplastin Time
  • Models, Animal
  • Materials Testing
  • Heparin
  • Heart-Assist Devices
  • Heart Valve Prosthesis Implantation
  • Fever
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Weiss, W. J., Carney, E. L., Clark, J. B., Peterson, R., Cooper, T. K., Nifong, T. P., … Pierce, W. S. (2012). Chronic in vivo testing of the Penn State infant ventricular assist device. ASAIO J, 58(1), 65–72. https://doi.org/10.1097/MAT.0b013e318239feb4
Weiss, William J., Elizabeth L. Carney, J Brian Clark, Rebecca Peterson, Timothy K. Cooper, Thomas P. Nifong, Christopher A. Siedlecki, et al. “Chronic in vivo testing of the Penn State infant ventricular assist device.ASAIO J 58, no. 1 (2012): 65–72. https://doi.org/10.1097/MAT.0b013e318239feb4.
Weiss WJ, Carney EL, Clark JB, Peterson R, Cooper TK, Nifong TP, et al. Chronic in vivo testing of the Penn State infant ventricular assist device. ASAIO J. 2012;58(1):65–72.
Weiss, William J., et al. “Chronic in vivo testing of the Penn State infant ventricular assist device.ASAIO J, vol. 58, no. 1, 2012, pp. 65–72. Pubmed, doi:10.1097/MAT.0b013e318239feb4.
Weiss WJ, Carney EL, Clark JB, Peterson R, Cooper TK, Nifong TP, Siedlecki CA, Hicks D, Doxtater B, Lukic B, Yeager E, Reibson J, Cysyk J, Rosenberg G, Pierce WS. Chronic in vivo testing of the Penn State infant ventricular assist device. ASAIO J. 2012;58(1):65–72.

Published In

ASAIO J

DOI

EISSN

1538-943X

Publication Date

2012

Volume

58

Issue

1

Start / End Page

65 / 72

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Sheep
  • Prosthesis Design
  • Partial Thromboplastin Time
  • Models, Animal
  • Materials Testing
  • Heparin
  • Heart-Assist Devices
  • Heart Valve Prosthesis Implantation
  • Fever