Skip to main content
Journal cover image

Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit.

Publication ,  Journal Article
Qiu, F; Clark, JB; Kunselman, AR; Undar, A; Myers, JL
Published in: Perfusion
July 2011

OBJECTIVE: To construct an ideal extracorporeal life support (ECLS) circuit in terms of hemodynamic performance, each component of the circuit should be evaluated. Most cannulae manufacturers evaluate their products using water as the priming solution. We conducted this study to evaluate the different sizes of arterial and venous cannulae in a simulated neonatal ECLS circuit primed with human blood. METHODS: The simulated neonatal ECLS circuit was composed of a Capiox Baby RX05 oxygenator, a Rotaflow centrifugal pump and a heater & cooler unit. Three Medtronic Bio-Medicus arterial cannulae (8Fr, 10Fr, 12Fr) and three venous cannulae (10Fr, 12Fr, 14Fr) were tested in seven combinations (8A-10V, 8A-12V, 10A-10V, 10A-12V, 10A-14V, 12A-12V, 12A-14V). All the experiments were conducted using human blood at a hematocrit of 40% and at a constant temperature of 37°C. The "tip to tip" priming volume of the entire circuit was 135ml. The blood volume of the pseudo patient was 500ml. RESULTS: Flow rates increased linearly with increasing size in both venous and arterial cannulae at the same pump speeds. The increase in flow rate was greater when changing the arterial cannulae (next size larger) compared to changing the venous cannulae (next size larger). The pressure drops of the arterial cannula were correlated with the flow rates, regardless of the pseudo patient pressure and the venous cannula used simultaneously. CONCLUSIONS: The results show the difference in flow ranges and pressure drops of seven combinations of arterial and venous cannulae. It also suggests that the arterial cannula, not the venous cannula, has greater impact on the flow rate when a centrifugal pump is used in a neonatal ECLS circuit. The results of this study have been translated to further advancing the clinical practice in our institution.

Duke Scholars

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

July 2011

Volume

26

Issue

4

Start / End Page

276 / 283

Location

England

Related Subject Headings

  • Veins
  • Infant, Newborn
  • Humans
  • Hemodynamics
  • Extracorporeal Circulation
  • Catheters
  • Catheterization
  • Arteries
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Qiu, F., Clark, J. B., Kunselman, A. R., Undar, A., & Myers, J. L. (2011). Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit. Perfusion, 26(4), 276–283. https://doi.org/10.1177/0267659111406768
Qiu, F., J. B. Clark, A. R. Kunselman, A. Undar, and J. L. Myers. “Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit.Perfusion 26, no. 4 (July 2011): 276–83. https://doi.org/10.1177/0267659111406768.
Qiu F, Clark JB, Kunselman AR, Undar A, Myers JL. Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit. Perfusion. 2011 Jul;26(4):276–83.
Qiu, F., et al. “Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit.Perfusion, vol. 26, no. 4, July 2011, pp. 276–83. Pubmed, doi:10.1177/0267659111406768.
Qiu F, Clark JB, Kunselman AR, Undar A, Myers JL. Hemodynamic evaluation of arterial and venous cannulae performance in a simulated neonatal extracorporeal life support circuit. Perfusion. 2011 Jul;26(4):276–283.
Journal cover image

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

July 2011

Volume

26

Issue

4

Start / End Page

276 / 283

Location

England

Related Subject Headings

  • Veins
  • Infant, Newborn
  • Humans
  • Hemodynamics
  • Extracorporeal Circulation
  • Catheters
  • Catheterization
  • Arteries
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology