An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model.

Publication ,  Journal Article
Clark, JB; Guan, Y; McCoach, R; Kunselman, AR; Myers, JL; Undar, A
Published in: Perfusion
May 2011

During extracorporeal life support with centrifugal blood pumps, retrograde pump flow may occur when the pump revolutions decrease below a critical value determined by the circuit resistance and the characteristics of the pump. We created a laboratory model to evaluate the occurrence of retrograde flow in each of three centrifugal blood pumps: the Rotaflow, the CentriMag, and the Bio-Medicus BP-50. At simulated patient pressures of 60, 80, and 100 mmHg, each pump was evaluated at speeds from 1000 to 2200 rpm and flow rates were measured. Retrograde flow occurred at low revolution speeds in all three centrifugal pumps. The Bio-Medicus pump was the least likely to demonstrate retrograde flow at low speeds, followed by the Rotaflow pump. The CentriMag pump showed the earliest transition to retrograde flow, as well as the highest degree of retrograde flow. At every pump speed evaluated, the Bio-Medicus pump delivered the highest antegrade flow and the CentriMag pump delivered the least.

Duke Scholars

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

May 2011

Volume

26

Issue

3

Start / End Page

185 / 190

Location

England

Related Subject Headings

  • Male
  • Humans
  • Heart-Assist Devices
  • Female
  • Extracorporeal Circulation
  • Child, Preschool
  • Child
  • Blood Flow Velocity
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Clark, J. B., Guan, Y., McCoach, R., Kunselman, A. R., Myers, J. L., & Undar, A. (2011). An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model. Perfusion, 26(3), 185–190. https://doi.org/10.1177/0267659110394712
Clark, Joseph B., Yulong Guan, Robert McCoach, Allen R. Kunselman, John L. Myers, and Akif Undar. “An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model.Perfusion 26, no. 3 (May 2011): 185–90. https://doi.org/10.1177/0267659110394712.
Clark, Joseph B., et al. “An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model.Perfusion, vol. 26, no. 3, May 2011, pp. 185–90. Pubmed, doi:10.1177/0267659110394712.
Journal cover image

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

May 2011

Volume

26

Issue

3

Start / End Page

185 / 190

Location

England

Related Subject Headings

  • Male
  • Humans
  • Heart-Assist Devices
  • Female
  • Extracorporeal Circulation
  • Child, Preschool
  • Child
  • Blood Flow Velocity
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology