Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit.

Publication ,  Journal Article
Oh, BH; Kim, JY; Lee, JB; Hong, JT; Sung, JH; Than, KD; Lee, HJ; Kim, IS
Published in: Neurospine
December 2023

OBJECTIVE: Achieving successful fusion during spine surgery is dependent on rigid pedicle screw fixation. To assess fixation strength, the insertional torque can be measured during intraoperative screw fixation. This study aimed to explore the technical feasibility of measuring the insertional torque of a pedicle screw, while investigating its relationship with bone density. METHODS: Thoraco-lumbar screw fixation fusion surgery was performed on 53 patients (mean age, 65.5 ± 9.8 years). The insertional torque of 284 screws was measured at the point passing through the pedicle using a calibrated torque wrench, with a specially designed connector to the spine screw system. The Hounsfield units (HU) value was determined by assessing the trabecular portion of the index vertebral body on sagittal computed tomography images. We analyzed the relationship between the measured insertional torque and the following bone strength parameters: bone mineral density (BMD) and HU of the vertebral body. RESULTS: The mean insertion torque was 105.55 ± 58.08 N∙cm and T-score value (BMD) was -1.14 ± 1.49. Mean HU value was 136.37 ± 57.59. Screw insertion torque was positively correlated with BMD and HU in whole patients. However, in cases of osteopenia, all variables showed very weak correlations with insertional torque. In patients with osteoporosis, there was no statistically significant correlation between BMD and torque strength; HU showed a significant correlation. CONCLUSION: The insertional torque of screw fixation significantly correlated with bone density (BMD and HU). HU measurements showed greater clinical significance than did BMD values in patients with osteoporosis.

Duke Scholars

Published In

Neurospine

DOI

ISSN

2586-6583

Publication Date

December 2023

Volume

20

Issue

4

Start / End Page

1177 / 1185

Location

Korea (South)
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Oh, B. H., Kim, J. Y., Lee, J. B., Hong, J. T., Sung, J. H., Than, K. D., … Kim, I. S. (2023). Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit. Neurospine, 20(4), 1177–1185. https://doi.org/10.14245/ns.2346830.415
Oh, Byeong Ho, Jee Yong Kim, Jong Beom Lee, Jae Taek Hong, Jae Hoon Sung, Khoi D. Than, Ho Jin Lee, and Il Sup Kim. “Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit.Neurospine 20, no. 4 (December 2023): 1177–85. https://doi.org/10.14245/ns.2346830.415.
Oh BH, Kim JY, Lee JB, Hong JT, Sung JH, Than KD, et al. Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit. Neurospine. 2023 Dec;20(4):1177–85.
Oh, Byeong Ho, et al. “Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit.Neurospine, vol. 20, no. 4, Dec. 2023, pp. 1177–85. Pubmed, doi:10.14245/ns.2346830.415.
Oh BH, Kim JY, Lee JB, Hong JT, Sung JH, Than KD, Lee HJ, Kim IS. Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit. Neurospine. 2023 Dec;20(4):1177–1185.

Published In

Neurospine

DOI

ISSN

2586-6583

Publication Date

December 2023

Volume

20

Issue

4

Start / End Page

1177 / 1185

Location

Korea (South)