Skip to main content

An unusual intrathecal baclofen pump failure thirteen months after implantation

Publication ,  Journal Article
Vorenkamp, KE; Gandhi, SH; Manolides, AS; Warren, DT
Published in: Interventional Pain Medicine
December 1, 2023

Objective: To report a case of intrathecal pump failure following months of diminishing benefit from intrathecal baclofen, and to heighten awareness that intrathecal pump malfunction can occur without precipitating events. Case report: A 40 year-old woman with multiple sclerosis and spastic paraplegia developed worsening spasticity after ten months of therapeutic stability with intrathecal baclofen. After other causes were pursued, this was discovered to be due to pump malfunction only thirteen months after implantation. After pump interrogation and discussion with the manufacturer the pump was replaced urgently and the patient regained therapeutic benefit and had no further complications. Conclusions: We present herein what we believe to be the first report of a verified pump malfunction resulting in disruption of intrathecal baclofen delivery within thirteen months of implantation. Due to the possible severity of acute baclofen withdrawal, the pump was replaced urgently after diagnosis. Because of the patient's and her healthcare providers' vigilance, she experienced no adverse events. Healthcare providers are encouraged to acknowledge the possibility of intrathecal pump malfunction in similar scenarios, ensuring patient safety while systematically examining the underlying problem.

Duke Scholars

Published In

Interventional Pain Medicine

DOI

ISSN

2772-5944

Publication Date

December 1, 2023

Volume

2

Issue

4
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Vorenkamp, K. E., Gandhi, S. H., Manolides, A. S., & Warren, D. T. (2023). An unusual intrathecal baclofen pump failure thirteen months after implantation. Interventional Pain Medicine, 2(4). https://doi.org/10.1016/j.inpm.2023.100284
Vorenkamp, K. E., S. H. Gandhi, A. S. Manolides, and D. T. Warren. “An unusual intrathecal baclofen pump failure thirteen months after implantation.” Interventional Pain Medicine 2, no. 4 (December 1, 2023). https://doi.org/10.1016/j.inpm.2023.100284.
Vorenkamp KE, Gandhi SH, Manolides AS, Warren DT. An unusual intrathecal baclofen pump failure thirteen months after implantation. Interventional Pain Medicine. 2023 Dec 1;2(4).
Vorenkamp, K. E., et al. “An unusual intrathecal baclofen pump failure thirteen months after implantation.” Interventional Pain Medicine, vol. 2, no. 4, Dec. 2023. Scopus, doi:10.1016/j.inpm.2023.100284.
Vorenkamp KE, Gandhi SH, Manolides AS, Warren DT. An unusual intrathecal baclofen pump failure thirteen months after implantation. Interventional Pain Medicine. 2023 Dec 1;2(4).

Published In

Interventional Pain Medicine

DOI

ISSN

2772-5944

Publication Date

December 1, 2023

Volume

2

Issue

4