Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom.

Publication ,  Journal Article
Jiang, Y; Ma, D; Keenan, KE; Stupic, KF; Gulani, V; Griswold, MA
Published in: Magn Reson Med
October 2017

PURPOSE: The purpose of this study was to evaluate accuracy and repeatability of T1 and T2 estimates of a MR fingerprinting (MRF) method using the ISMRM/NIST MRI system phantom. METHODS: The ISMRM/NIST MRI system phantom contains multiple compartments with standardized T1 , T2 , and proton density values. Conventional inversion-recovery spin echo and spin echo methods were used to characterize the T1 and T2 values in the phantom. The phantom was scanned using the MRF-FISP method over 34 consecutive days. The mean T1 and T2 values were compared with the values from the spin echo methods. The repeatability was characterized as the coefficient of variation of the measurements over 34 days. RESULTS: T1 and T2 values from MRF-FISP over 34 days showed a strong linear correlation with the measurements from the spin echo methods (R2  = 0.999 for T1 ; R2  = 0.996 for T2 ). The MRF estimates over the wide ranges of T1 and T2 values have less than 5% variation, except for the shortest T2 relaxation times where the method still maintains less than 8% variation. CONCLUSION: MRF measurements of T1 and T2 are highly repeatable over time and across wide ranges of T1 and T2 values. Magn Reson Med 78:1452-1457, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Duke Scholars

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

October 2017

Volume

78

Issue

4

Start / End Page

1452 / 1457

Location

United States

Related Subject Headings

  • Phantoms, Imaging
  • Pattern Recognition, Automated
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Jiang, Y., Ma, D., Keenan, K. E., Stupic, K. F., Gulani, V., & Griswold, M. A. (2017). Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn Reson Med, 78(4), 1452–1457. https://doi.org/10.1002/mrm.26509
Jiang, Yun, Dan Ma, Kathryn E. Keenan, Karl F. Stupic, Vikas Gulani, and Mark A. Griswold. “Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom.Magn Reson Med 78, no. 4 (October 2017): 1452–57. https://doi.org/10.1002/mrm.26509.
Jiang Y, Ma D, Keenan KE, Stupic KF, Gulani V, Griswold MA. Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn Reson Med. 2017 Oct;78(4):1452–7.
Jiang, Yun, et al. “Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom.Magn Reson Med, vol. 78, no. 4, Oct. 2017, pp. 1452–57. Pubmed, doi:10.1002/mrm.26509.
Jiang Y, Ma D, Keenan KE, Stupic KF, Gulani V, Griswold MA. Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn Reson Med. 2017 Oct;78(4):1452–1457.
Journal cover image

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

October 2017

Volume

78

Issue

4

Start / End Page

1452 / 1457

Location

United States

Related Subject Headings

  • Phantoms, Imaging
  • Pattern Recognition, Automated
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering