Skip to main content
Journal cover image

Slice profile and B1 corrections in 2D magnetic resonance fingerprinting.

Publication ,  Journal Article
Ma, D; Coppo, S; Chen, Y; McGivney, DF; Jiang, Y; Pahwa, S; Gulani, V; Griswold, MA
Published in: Magn Reson Med
November 2017

PURPOSE: The goal of this study is to characterize and improve the accuracy of 2D magnetic resonance fingerprinting (MRF) scans in the presence of slice profile (SP) and B1 imperfections, which are two main factors that affect quantitative results in MRF. METHODS: The SP and B1 imperfections are characterized and corrected separately. The SP effect is corrected by simulating the radiofrequency pulse in the dictionary, and the B1 is corrected by acquiring a B1 map using the Bloch-Siegert method before each scan. The accuracy, precision, and repeatability of the proposed method are evaluated in phantom studies. The effects of both SP and B1 imperfections are also illustrated and corrected in the in vivo studies. RESULTS: The SP and B1 corrections improve the accuracy of the T1 and T2 values, independent of the shape of the radiofrequency pulse. The T1 and T2 values obtained from different excitation patterns become more consistent after corrections, which leads to an improvement of the robustness of the MRF design. CONCLUSION: This study demonstrates that MRF is sensitive to both SP and B1 effects, and that corrections can be made to improve the accuracy of MRF with only a 2-s increase in acquisition time. Magn Reson Med 78:1781-1789, 2017. © 2017 International Society for Magnetic Resonance in Medicine.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

November 2017

Volume

78

Issue

5

Start / End Page

1781 / 1789

Location

United States

Related Subject Headings

  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • Algorithms
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Ma, D., Coppo, S., Chen, Y., McGivney, D. F., Jiang, Y., Pahwa, S., … Griswold, M. A. (2017). Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med, 78(5), 1781–1789. https://doi.org/10.1002/mrm.26580
Ma, Dan, Simone Coppo, Yong Chen, Debra F. McGivney, Yun Jiang, Shivani Pahwa, Vikas Gulani, and Mark A. Griswold. “Slice profile and B1 corrections in 2D magnetic resonance fingerprinting.Magn Reson Med 78, no. 5 (November 2017): 1781–89. https://doi.org/10.1002/mrm.26580.
Ma D, Coppo S, Chen Y, McGivney DF, Jiang Y, Pahwa S, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017 Nov;78(5):1781–9.
Ma, Dan, et al. “Slice profile and B1 corrections in 2D magnetic resonance fingerprinting.Magn Reson Med, vol. 78, no. 5, Nov. 2017, pp. 1781–89. Pubmed, doi:10.1002/mrm.26580.
Ma D, Coppo S, Chen Y, McGivney DF, Jiang Y, Pahwa S, Gulani V, Griswold MA. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017 Nov;78(5):1781–1789.
Journal cover image

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

November 2017

Volume

78

Issue

5

Start / End Page

1781 / 1789

Location

United States

Related Subject Headings

  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • Algorithms
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering