Skip to main content
Journal cover image

Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling.

Journal articles  - Journal Article
White, R; Sarathy, S; Badheka, A; Turek, JW; Raghavan, ML; Chivukula, VK
Published in: Perfusion
September 2026

IntroductionCurrent extracorporeal membrane oxygenation (ECMO) monitoring methods for circuit obstructions have limitations. In this study, we test the hypothesis that changes in flow in a pre-existing shunt tubing of the ECMO circuit can be used as a predictor of obstructions for both venovenous (V-V) and venoarterial ECMO (V-A ECMO) circuits using a computational hemodynamic model.MethodsA custom computational hemodynamic model (CHM) representing the blood flow within the V-A ECMO and V-V ECMO circuit and native cardiovascular system was developed for a hemodynamic parametric analysis. Several degrees of obstructions within the oxygenator and outlet cannula were simulated, and the flow in the ECMO circuit and shunt was monitored to assess its sensitivity to circuit obstructions. Over 500 cases were analyzed to investigate various clinically-relevant levels of ECMO support (2 - 5.5 L/min).ResultsOur model demonstrated that shunt flow changed linearly with increasing oxygenator obstruction for all oxygenator obstruction scenarios tested, which matched experimental data. Similar relationships were also seen for all outflow cannula obstructions.ConclusionsThe findings are consistent with our hypothesis, indicating the sensitivity of the shunt flow rate as a predictor of ECMO circuit obstructions for both V-A and V-V ECMO circuits. The data supports that shunt flow responds to hemodynamic changes throughout the ECMO-patient circuit and thus can be used as an early indicator for ECMO system obstructions regardless of V-A or V-V ECMO setup, indicating its robustness as an early warning indicator of ECMO system obstructions.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

September 2026

Volume

41

Issue

6

Start / End Page

917 / 927

Location

England

Related Subject Headings

  • Models, Cardiovascular
  • Humans
  • Hemodynamics
  • Extracorporeal Membrane Oxygenation
  • 3201 Cardiovascular medicine and haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
White, R., Sarathy, S., Badheka, A., Turek, J. W., Raghavan, M. L., & Chivukula, V. K. (2026). Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling. Perfusion, 41(6), 917–927. https://doi.org/10.1177/02676591251389786
White, Ruth, Srivats Sarathy, Aditya Badheka, Joseph W. Turek, Madhavan L. Raghavan, and Venkat Keshav Chivukula. “Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling.” Perfusion 41, no. 6 (September 2026): 917–27. https://doi.org/10.1177/02676591251389786.
White R, Sarathy S, Badheka A, Turek JW, Raghavan ML, Chivukula VK. Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling. Perfusion. 2026 Sep;41(6):917–27.
White, Ruth, et al. “Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling.” Perfusion, vol. 41, no. 6, Sept. 2026, pp. 917–27. Pubmed, doi:10.1177/02676591251389786.
White R, Sarathy S, Badheka A, Turek JW, Raghavan ML, Chivukula VK. Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling. Perfusion. 2026 Sep;41(6):917–927.
Journal cover image

Published In

Perfusion

DOI

EISSN

1477-111X

Publication Date

September 2026

Volume

41

Issue

6

Start / End Page

917 / 927

Location

England

Related Subject Headings

  • Models, Cardiovascular
  • Humans
  • Hemodynamics
  • Extracorporeal Membrane Oxygenation
  • 3201 Cardiovascular medicine and haematology