A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair

Publication ,  Journal Article
Wang, L; Zhou, X; Guo, D; Hou, K; Shi, Z; Tang, X; Fu, W
Published in: Journal of Endovascular Therapy
August 2018

Purpose: To describe a new adjustable puncture system for in situ fenestration in thoracic endovascular aortic repair (TEVAR). Technique: An adjustable puncture needle for use in conjunction with a steerable 8-F, 55-cm Fustar sheath is demonstrated in a 65-year-old man with acute complicated type B dissection involving the left subclavian artery (LSA). The puncture device features an inflatable balloon at the tip, a central lumen for 0.018-inch guidewires, and a 3-level puncture depth. After thoracic stent-graft deployment at zone 2, the needle/sheath combination was delivered from a left brachial artery access. The needle was adjusted perpendicular to the fabric of the stent-graft with the assistance of the steerable sheath. The balloon at the tip was inflated to center the needle, and the puncture depth was selected on the puncture needle system. Holding the sheath and puncture needle together, a hole was created in the graft fabric. The aperture was sequentially dilated to accommodate the mating stent selected to maintain perfusion to the LSA. This new device has been successfully applied in 6 patients treated with TEVAR for different arch pathologies. Conclusion: This new puncture device could assist in situ fenestration and improve the technical success rate.

Duke Scholars

Published In

Journal of Endovascular Therapy

DOI

EISSN

1545-1550

ISSN

1526-6028

Publication Date

August 2018

Volume

25

Issue

4

Start / End Page

474 / 479

Publisher

SAGE Publications

Related Subject Headings

  • Cardiovascular System & Hematology
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Wang, L., Zhou, X., Guo, D., Hou, K., Shi, Z., Tang, X., & Fu, W. (2018). A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair. Journal of Endovascular Therapy, 25(4), 474–479. https://doi.org/10.1177/1526602818776623
Wang, Lixin, Xiushi Zhou, Daqiao Guo, Kai Hou, Zhenyu Shi, Xiao Tang, and Weiguo Fu. “A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair.” Journal of Endovascular Therapy 25, no. 4 (August 2018): 474–79. https://doi.org/10.1177/1526602818776623.
Wang L, Zhou X, Guo D, Hou K, Shi Z, Tang X, et al. A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair. Journal of Endovascular Therapy. 2018 Aug;25(4):474–9.
Wang, Lixin, et al. “A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair.” Journal of Endovascular Therapy, vol. 25, no. 4, SAGE Publications, Aug. 2018, pp. 474–79. Crossref, doi:10.1177/1526602818776623.
Wang L, Zhou X, Guo D, Hou K, Shi Z, Tang X, Fu W. A New Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair. Journal of Endovascular Therapy. SAGE Publications; 2018 Aug;25(4):474–479.
Journal cover image

Published In

Journal of Endovascular Therapy

DOI

EISSN

1545-1550

ISSN

1526-6028

Publication Date

August 2018

Volume

25

Issue

4

Start / End Page

474 / 479

Publisher

SAGE Publications

Related Subject Headings

  • Cardiovascular System & Hematology
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology
  • 1103 Clinical Sciences