The goal of tissue characterization via quantitative computed tomography can only be accomplished with careful control of certain parameters. A quality assurance program is described which includes daily measurement of certain essential elements e.g. Hounsfield conversion, linearity, noise and resolution. Secondary measurements of collimation, bed movement, dose, and sensitivity profiles are performed on a monthly basis. Daily measurements are accomplished in less than 15 minutes while monthly checks take less than 1.5 hr thus minimizing conflict with a busy clinical routine. Examples of rapid recording methods and measurement techniques are included. © 1982 SPIE.
Duke Scholars
Published In
Proceedings of SPIE the International Society for Optical Engineering
DOI
EISSN
1996-756X
ISSN
0277-786X
Publication Date
December 29, 1982
Volume
347
Start / End Page
148 / 154
Related Subject Headings
- 5102 Atomic, molecular and optical physics
- 4009 Electronics, sensors and digital hardware
- 4006 Communications engineering
Citation
APA
Chicago
ICMJE
MLA
NLM
Johnson, G. A. (1982). Practical approach to quality assurance in computed tomography. Proceedings of SPIE the International Society for Optical Engineering, 347, 148–154. https://doi.org/10.1117/12.933821
Johnson, G. A. “Practical approach to quality assurance in computed tomography.” Proceedings of SPIE the International Society for Optical Engineering 347 (December 29, 1982): 148–54. https://doi.org/10.1117/12.933821.
Johnson GA. Practical approach to quality assurance in computed tomography. Proceedings of SPIE the International Society for Optical Engineering. 1982 Dec 29;347:148–54.
Johnson, G. A. “Practical approach to quality assurance in computed tomography.” Proceedings of SPIE the International Society for Optical Engineering, vol. 347, Dec. 1982, pp. 148–54. Scopus, doi:10.1117/12.933821.
Johnson GA. Practical approach to quality assurance in computed tomography. Proceedings of SPIE the International Society for Optical Engineering. 1982 Dec 29;347:148–154.
Published In
Proceedings of SPIE the International Society for Optical Engineering
DOI
EISSN
1996-756X
ISSN
0277-786X
Publication Date
December 29, 1982
Volume
347
Start / End Page
148 / 154
Related Subject Headings
- 5102 Atomic, molecular and optical physics
- 4009 Electronics, sensors and digital hardware
- 4006 Communications engineering