MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging.
Publication
, Journal Article
Hedlund, LW; Cofer, GP; Owen, SJ; Allan Johnson, G
Published in: Magn Reson Imaging
July 2000
We describe an MR-compatible ventilator that is computer controlled to generate a variety of breathing patterns, to minimize image degrading effects of breathing motion, and to support delivery of gas anesthesia and experimental inhalational gases. A key feature of this ventilator is the breathing valve that attaches directly to the endotracheal tube to reduce dead volume and allows independent control of inspiratory and expiratory phases of ventilation. This ventilator has been used in a wide variety of MR and x-ray microscopy studies of small animals, especially for MR imaging the lungs with hyperpolarized gases ((3)He & (129)Xe).
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
Magn Reson Imaging
DOI
ISSN
0730-725X
Publication Date
July 2000
Volume
18
Issue
6
Start / End Page
753 / 759
Location
Netherlands
Related Subject Headings
- Ventilators, Mechanical
- Ventilation
- Rodentia
- Rats
- Protons
- Nuclear Medicine & Medical Imaging
- Noble Gases
- Mice
- Magnetic Resonance Imaging
- Lung
Citation
APA
Chicago
ICMJE
MLA
NLM
Hedlund, L. W., Cofer, G. P., Owen, S. J., & Allan Johnson, G. (2000). MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging. Magn Reson Imaging, 18(6), 753–759. https://doi.org/10.1016/s0730-725x(00)00154-5
Hedlund, L. W., G. P. Cofer, S. J. Owen, and G. Allan Johnson. “MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging.” Magn Reson Imaging 18, no. 6 (July 2000): 753–59. https://doi.org/10.1016/s0730-725x(00)00154-5.
Hedlund LW, Cofer GP, Owen SJ, Allan Johnson G. MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging. Magn Reson Imaging. 2000 Jul;18(6):753–9.
Hedlund, L. W., et al. “MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging.” Magn Reson Imaging, vol. 18, no. 6, July 2000, pp. 753–59. Pubmed, doi:10.1016/s0730-725x(00)00154-5.
Hedlund LW, Cofer GP, Owen SJ, Allan Johnson G. MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging. Magn Reson Imaging. 2000 Jul;18(6):753–759.
Published In
Magn Reson Imaging
DOI
ISSN
0730-725X
Publication Date
July 2000
Volume
18
Issue
6
Start / End Page
753 / 759
Location
Netherlands
Related Subject Headings
- Ventilators, Mechanical
- Ventilation
- Rodentia
- Rats
- Protons
- Nuclear Medicine & Medical Imaging
- Noble Gases
- Mice
- Magnetic Resonance Imaging
- Lung