Plate scatter correction for improved performance in dual-energy imaging.

Publication ,  Journal Article
Hinshaw, DA; Dobbins, JT
Published in: Med Phys
June 1996

We have developed a method for correcting for the scatter produced by the computed radiography plates in single-shot dual-energy imaging. In our technique, which uses a series of four computed radiography plates (the middle two plates act as a filter), the back plate can have scatter fractions as high as 30% due to the scatter from the first three plates. We use the deconvolution method to correct for this scatter, as the scatter medium is uniform and isotropic with a constant air gap. The ratio of the modulation transfer function (MTF) of a standard computed radiography plate to that of a dual-energy plate gives us the correction factor we need to scale the Fourier transforms of our images, thus compensating for the reduced response of the system due to scatter. We also investigated the effect of having an air gap between the back dual-energy plate and the first three plates. Our measurements show that this degrades the signal-to-noise response of the back plate slightly at low spatial frequencies. We also used independent scatter fraction measurements to verify the validity of our MTF measurements. Applying the correction to a back plate image improves the fine detail contrast by approximately 30%.

Duke Scholars

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

June 1996

Volume

23

Issue

6

Start / End Page

871 / 876

Location

United States

Related Subject Headings

  • Scattering, Radiation
  • Radiography, Dual-Energy Scanned Projection
  • Nuclear Medicine & Medical Imaging
  • Models, Theoretical
  • Humans
  • Biophysics
  • Biophysical Phenomena
  • Air
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Hinshaw, D. A., & Dobbins, J. T. (1996). Plate scatter correction for improved performance in dual-energy imaging. Med Phys, 23(6), 871–876. https://doi.org/10.1118/1.597855
Hinshaw, D. A., and J. T. Dobbins. “Plate scatter correction for improved performance in dual-energy imaging.Med Phys 23, no. 6 (June 1996): 871–76. https://doi.org/10.1118/1.597855.
Hinshaw DA, Dobbins JT. Plate scatter correction for improved performance in dual-energy imaging. Med Phys. 1996 Jun;23(6):871–6.
Hinshaw, D. A., and J. T. Dobbins. “Plate scatter correction for improved performance in dual-energy imaging.Med Phys, vol. 23, no. 6, June 1996, pp. 871–76. Pubmed, doi:10.1118/1.597855.
Hinshaw DA, Dobbins JT. Plate scatter correction for improved performance in dual-energy imaging. Med Phys. 1996 Jun;23(6):871–876.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

June 1996

Volume

23

Issue

6

Start / End Page

871 / 876

Location

United States

Related Subject Headings

  • Scattering, Radiation
  • Radiography, Dual-Energy Scanned Projection
  • Nuclear Medicine & Medical Imaging
  • Models, Theoretical
  • Humans
  • Biophysics
  • Biophysical Phenomena
  • Air
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering