Characteristics of sensitometric curves of radiographic films.

Publication ,  Journal Article
Zhu, XR; Yoo, S; Jursinic, PA; Grimm, DF; Lopez, F; Rownd, JJ; Gillin, MT
Published in: Med Phys
May 2003

A new type of radiographic film, EDR (extended dose range) film, has been recently become available for film dosimetry. It is particularly attractive for composite isodose verification of intensity modulated radiation therapy because of its low sensitivity relative to the more common Kodak XV film. For XV film, the relationship between optical density and dose, commonly known as the sensitometric curve, depends linearly on the dose at low densities. Unlike XV film, the sensitometric curve of EDR film irradiated by megavoltage x rays is not linearly dependent on the dose at low densities. In this work, to understand the mechanisms governing the shape of the sensitometric curves, EDR film was studied with kilovoltage x rays, 60Co gamma rays, megavoltage x rays, and electron beams. As a comparison, XV film was also studied with the same beams mentioned above. The model originally developed by Silberstein [J. Opt. Soc. Am. 35, 93-107, 1945)] is used to fit experimental data. It is found that the single hit model can be used to predict the sensitometric curve for XV films irradiated by all beams used in this work and for EDR films exposed to kilovoltage x rays. For EDR film irradiated by 60Co gamma rays, megavoltage x rays, and electron beams, the double hit model is used to fit the sensitometric curves. For doses less than 100 cGy, a systematic difference between measured densities and that predicted by the double hit model is observed. Possible causes of the observed differences are discussed. The results of this work provide a theoretical explanation of the sensitometric behavior of EDR film.

Duke Scholars

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

May 2003

Volume

30

Issue

5

Start / End Page

912 / 919

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Reproducibility of Results
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Film Dosimetry
  • Equipment Failure Analysis
  • Equipment Design
  • Dose-Response Relationship, Radiation
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Zhu, X. R., Yoo, S., Jursinic, P. A., Grimm, D. F., Lopez, F., Rownd, J. J., & Gillin, M. T. (2003). Characteristics of sensitometric curves of radiographic films. Med Phys, 30(5), 912–919. https://doi.org/10.1118/1.1568979
Zhu, X. R., S. Yoo, P. A. Jursinic, D. F. Grimm, F. Lopez, J. J. Rownd, and M. T. Gillin. “Characteristics of sensitometric curves of radiographic films.Med Phys 30, no. 5 (May 2003): 912–19. https://doi.org/10.1118/1.1568979.
Zhu XR, Yoo S, Jursinic PA, Grimm DF, Lopez F, Rownd JJ, et al. Characteristics of sensitometric curves of radiographic films. Med Phys. 2003 May;30(5):912–9.
Zhu, X. R., et al. “Characteristics of sensitometric curves of radiographic films.Med Phys, vol. 30, no. 5, May 2003, pp. 912–19. Pubmed, doi:10.1118/1.1568979.
Zhu XR, Yoo S, Jursinic PA, Grimm DF, Lopez F, Rownd JJ, Gillin MT. Characteristics of sensitometric curves of radiographic films. Med Phys. 2003 May;30(5):912–919.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

May 2003

Volume

30

Issue

5

Start / End Page

912 / 919

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Reproducibility of Results
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Film Dosimetry
  • Equipment Failure Analysis
  • Equipment Design
  • Dose-Response Relationship, Radiation
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering