Publication ,  Journal Article
Frush, DP; Yoshizumi, T
Published in: Pediatr Radiol
September 2006

Tremendous advances have been made in imaging in children with both congenital and acquired heart disease. These include technical advances in cardiac catheterization and conventional angiography, especially with advancements in interventional procedures, as well as noninvasive imaging with MR and CT angiography. With rapid advances in multidetector CT (MDCT) technology, most recently 64-detector array systems (64-slice MDCT), have come a number of advantages over MR. However, both conventional and CT angiography impart radiation dose to children. Although the presence of radiation exposure to children has long been recognized, it is apparent that our ability to assess this dose, particularly in light of the rapid advancements, has been limited. Traditional methods of dosimetry for both conventional and CT angiography are somewhat cumbersome or involve a potential for substantial uncertainty. Recent developments in dosimetry, including metal oxide semiconductor field effect transistors (MOSFET) and the availability of anthropomorphic, tissue-equivalent phantoms have provided new opportunities for dosimetric assessments. Recent work with this technology in state-of-the-art cardiac angiography suites as well as with MDCT have offered direct comparisons of doses in infants and children undergoing diagnostic cardiac evaluation. It is with these dose data that assessment of risks, and ultimately the assessment of risk-benefit, can be better achieved.

Duke Scholars

Published In

Pediatr Radiol

DOI

EISSN

1432-1998

Publication Date

September 2006

Volume

36 Suppl 2

Issue

Suppl 2

Start / End Page

154 / 158

Location

Germany

Related Subject Headings

  • Tomography, X-Ray Computed
  • Relative Biological Effectiveness
  • Radiometry
  • Radiography, Interventional
  • Radiation Protection
  • Nuclear Medicine & Medical Imaging
  • Humans
  • Child
  • Body Burden
  • Angiography
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Frush, D. P., & Yoshizumi, T. (2006). Conventional and CT angiography in children: dosimetry and dose comparisons. Pediatr Radiol, 36 Suppl 2(Suppl 2), 154–158. https://doi.org/10.1007/s00247-006-0190-6
Frush, Donald P., and Terry Yoshizumi. “Conventional and CT angiography in children: dosimetry and dose comparisons.Pediatr Radiol 36 Suppl 2, no. Suppl 2 (September 2006): 154–58. https://doi.org/10.1007/s00247-006-0190-6.
Frush DP, Yoshizumi T. Conventional and CT angiography in children: dosimetry and dose comparisons. Pediatr Radiol. 2006 Sep;36 Suppl 2(Suppl 2):154–8.
Frush, Donald P., and Terry Yoshizumi. “Conventional and CT angiography in children: dosimetry and dose comparisons.Pediatr Radiol, vol. 36 Suppl 2, no. Suppl 2, Sept. 2006, pp. 154–58. Pubmed, doi:10.1007/s00247-006-0190-6.
Frush DP, Yoshizumi T. Conventional and CT angiography in children: dosimetry and dose comparisons. Pediatr Radiol. 2006 Sep;36 Suppl 2(Suppl 2):154–158.
Journal cover image

Published In

Pediatr Radiol

DOI

EISSN

1432-1998

Publication Date

September 2006

Volume

36 Suppl 2

Issue

Suppl 2

Start / End Page

154 / 158

Location

Germany

Related Subject Headings

  • Tomography, X-Ray Computed
  • Relative Biological Effectiveness
  • Radiometry
  • Radiography, Interventional
  • Radiation Protection
  • Nuclear Medicine & Medical Imaging
  • Humans
  • Child
  • Body Burden
  • Angiography