MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children.

Publication ,  Journal Article
Mukundan, S; Wang, PI; Frush, DP; Yoshizumi, T; Marcus, J; Kloeblen, E; Moore, M
Published in: AJR Am J Roentgenol
June 2007

OBJECTIVE: The purpose of our study was to measure radiation dose to the orbit during pediatric cranial CT with and without bismuth shielding using a novel dosimetry system. Cranial CT was performed on a pediatric anthropomorphic phantom, with and without bismuth eye shields. A solid-state metal oxide semiconductor field effect transistor (MOSFET) dosimeter was used to obtain real-time dose measurements. CONCLUSION: Bismuth shielding reduced radiation dose to the eye by up to 42%; shield artifact fell outside the diagnostic area of interest.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

AJR Am J Roentgenol

DOI

EISSN

1546-3141

Publication Date

June 2007

Volume

188

Issue

6

Start / End Page

1648 / 1650

Location

United States

Related Subject Headings

  • Transistors, Electronic
  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Reproducibility of Results
  • Relative Biological Effectiveness
  • Radiometry
  • Radiation Protection
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Humans
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Mukundan, S., Wang, P. I., Frush, D. P., Yoshizumi, T., Marcus, J., Kloeblen, E., & Moore, M. (2007). MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR Am J Roentgenol, 188(6), 1648–1650. https://doi.org/10.2214/AJR.06.1146
Mukundan, Srinivasan, Page Inman Wang, Donald P. Frush, Terry Yoshizumi, Jeffrey Marcus, Emily Kloeblen, and Meredith Moore. “MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children.AJR Am J Roentgenol 188, no. 6 (June 2007): 1648–50. https://doi.org/10.2214/AJR.06.1146.
Mukundan S, Wang PI, Frush DP, Yoshizumi T, Marcus J, Kloeblen E, et al. MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR Am J Roentgenol. 2007 Jun;188(6):1648–50.
Mukundan, Srinivasan, et al. “MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children.AJR Am J Roentgenol, vol. 188, no. 6, June 2007, pp. 1648–50. Pubmed, doi:10.2214/AJR.06.1146.
Mukundan S, Wang PI, Frush DP, Yoshizumi T, Marcus J, Kloeblen E, Moore M. MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR Am J Roentgenol. 2007 Jun;188(6):1648–1650.

Published In

AJR Am J Roentgenol

DOI

EISSN

1546-3141

Publication Date

June 2007

Volume

188

Issue

6

Start / End Page

1648 / 1650

Location

United States

Related Subject Headings

  • Transistors, Electronic
  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Reproducibility of Results
  • Relative Biological Effectiveness
  • Radiometry
  • Radiation Protection
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Humans