Feasibility of reversible pulmonary artery banding: early results and intermediate-term follow-up.
Development of a reversible pulmonary artery band might obviate the need for a second cardiac surgical procedure in children with some forms of congenital heart disease. We evaluated a segmented Silastic-coated Dacron mesh band 2 to 4 mm wide sewn together with absorbable 2-0 polydioxanone suture for use as a reversible pulmonary artery band. Nine puppies 6 to 8 weeks old (mean weight, 5.8 kg) underwent placement of this pulmonary artery band. All survived the operation with a mean initial systolic gradient of 6.5 +/- 1.6 mm Hg and underwent cardiac catheterization at monthly intervals for 3 months. Two puppies died of right ventricular failure. The remaining puppies underwent balloon angioplasty with balloons 20 to 25 mm in diameter. No complications resulted from balloon angioplasty, but 1 puppy died 24 hours later of unidentified causes. Balloon angioplasty decreased the mean gradient from 46.7 +/- 6.8 mm Hg to 6.7 +/- 2.6 mm Hg. Angiography showed an increase in mean diameter of the site of the pulmonary artery band from 5.2 +/- 1.0 to 10.8 +/- 1.7 mm Hg. There was no evidence of vessel injury on angiograms. Fourteen months after balloon angioplasty, the mean gradient was 22.3 +/- 17.0 mm Hg. Our data demonstrate that a functionally reversible pulmonary artery band constructed of segmented Silastic-coated Dacron mesh and 2-0 polydioxanone suture is feasible.
Duke Scholars
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Systole
- Surgical Mesh
- Silicone Elastomers
- Respiratory System
- Pulmonary Artery
- Polyethylene Terephthalates
- Heart Ventricles
- Dogs
- Cardiac Catheterization
- Blood Pressure
Citation
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Systole
- Surgical Mesh
- Silicone Elastomers
- Respiratory System
- Pulmonary Artery
- Polyethylene Terephthalates
- Heart Ventricles
- Dogs
- Cardiac Catheterization
- Blood Pressure