Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT.

Publication ,  Journal Article
Heneghan, JP; Spielmann, AL; Sheafor, DH; Kliewer, MA; DeLong, DM; Nelson, RC
Published in: J Comput Assist Tomogr
2002

PURPOSE: The purpose of this work was to compare the extent of pseudoenhancement (artifactual increase in measured attenuation of a simple cyst after contrast medium administration) in a phantom model on single detector and multidetector helical CT scanners. METHOD: The phantom consisted of four water-filled spheres varying in size from 8 to 28 mm, suspended in an aqueous contrast medium bath. Iodine concentration in the bath was varied: 0, 6, 12, and 24 mg/ml corresponding to attenuation values of 0, +108, +180, and +300 HU. The phantom was scanned on single detector and multidetector helical CT scanners during the same session. Collimation (1, 3, and 5 mm) and pitch (1 and 1.5:1, single detector; 3:1 and 6:1, multidetector) were varied at each concentration. All scans were performed at 140 kVp and 170 mA. The region of interest was measured at the center of each sphere. The effects were analyzed using a linear regression model. RESULTS: The degree of pseudoenhancement was more pronounced with increasing iodine concentration, decreasing cyst size, and wider collimation (all p = 0.0001). Pseudoenhancement was also more marked on the multidetector than the single detector scanner (p = 0.0001). At physiological levels of renal enhancement, the average pseudoenhancement was +18 HU for the single detector versus +23 HU for the multidetector scanner. Variation in pitch had no effect. CONCLUSION: Pseudoenhancement is greater on a multidetector than a single detector helical CT scanner and may exceed 20 HU at physiological levels of renal enhancement.

Duke Scholars

Published In

J Comput Assist Tomogr

DOI

ISSN

0363-8715

Publication Date

2002

Volume

26

Issue

1

Start / End Page

90 / 94

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Kidney Diseases, Cystic
  • Image Enhancement
  • Humans
  • Contrast Media
  • 4603 Computer vision and multimedia computation
  • 3202 Clinical sciences
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Heneghan, J. P., Spielmann, A. L., Sheafor, D. H., Kliewer, M. A., DeLong, D. M., & Nelson, R. C. (2002). Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT. J Comput Assist Tomogr, 26(1), 90–94. https://doi.org/10.1097/00004728-200201000-00013
Heneghan, Joan P., Audrey L. Spielmann, Douglas H. Sheafor, Mark A. Kliewer, David M. DeLong, and Rendon C. Nelson. “Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT.J Comput Assist Tomogr 26, no. 1 (2002): 90–94. https://doi.org/10.1097/00004728-200201000-00013.
Heneghan JP, Spielmann AL, Sheafor DH, Kliewer MA, DeLong DM, Nelson RC. Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT. J Comput Assist Tomogr. 2002;26(1):90–4.
Heneghan, Joan P., et al. “Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT.J Comput Assist Tomogr, vol. 26, no. 1, 2002, pp. 90–94. Pubmed, doi:10.1097/00004728-200201000-00013.
Heneghan JP, Spielmann AL, Sheafor DH, Kliewer MA, DeLong DM, Nelson RC. Pseudoenhancement of simple renal cysts: a comparison of single and multidetector helical CT. J Comput Assist Tomogr. 2002;26(1):90–94.

Published In

J Comput Assist Tomogr

DOI

ISSN

0363-8715

Publication Date

2002

Volume

26

Issue

1

Start / End Page

90 / 94

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Kidney Diseases, Cystic
  • Image Enhancement
  • Humans
  • Contrast Media
  • 4603 Computer vision and multimedia computation
  • 3202 Clinical sciences
  • 1103 Clinical Sciences