Extravasation detection accessory: clinical evaluation in 500 patients.

Publication ,  Journal Article
Birnbaum, BA; Nelson, RC; Chezmar, JL; Glick, SN
Published in: Radiology
August 1999

PURPOSE: To assess the ability of an extravasation detection accessory (EDA) to detect clinically important (> or = 10 mL) extravascular injection of iodinated contrast material delivered with an automated power injector. MATERIALS AND METHODS: Five hundred patients referred for contrast material-enhanced body computed tomography (CT) participated in a prospective, multi-institutional, observational study in which the EDA was used to identify and interrupt any injection associated with clinically important extravasation. The presence or absence of extravasation was definitively established with helical CT at the injection site (injection rate, from 1.0 to 5.0 mL/sec; mean, 2.9 mL/sec; median, 3.0 mL/sec). RESULTS: There were four true-positive (extravasation volumes, 13-18 mL), 484 true-negative, 12 false-positive, and no false-negative cases. The prevalence of overall and clinically important (> or = 10 mL) extravasation was 3.6% (18 of 500 cases) and 0.8% (four of 500 cases), respectively. The EDA had a sensitivity of 100% (four of four cases; 95% CI: 51%, 100%) and a specificity of 98% (484 of 496 cases; 95% CI: 96%, 99%) in the detection of clinically important extravasation. CONCLUSION: The EDA is easy to use, safe, and accurate in the monitoring of intravenous injections for extravasation, which could prove especially useful in high-flow-rate CT applications.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Radiology

DOI

ISSN

0033-8419

Publication Date

August 1999

Volume

212

Issue

2

Start / End Page

431 / 438

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Prospective Studies
  • Prevalence
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Injections, Intravenous
  • Humans
  • Female
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Birnbaum, B. A., Nelson, R. C., Chezmar, J. L., & Glick, S. N. (1999). Extravasation detection accessory: clinical evaluation in 500 patients. Radiology, 212(2), 431–438. https://doi.org/10.1148/radiology.212.2.r99au14431
Birnbaum, B. A., R. C. Nelson, J. L. Chezmar, and S. N. Glick. “Extravasation detection accessory: clinical evaluation in 500 patients.Radiology 212, no. 2 (August 1999): 431–38. https://doi.org/10.1148/radiology.212.2.r99au14431.
Birnbaum BA, Nelson RC, Chezmar JL, Glick SN. Extravasation detection accessory: clinical evaluation in 500 patients. Radiology. 1999 Aug;212(2):431–8.
Birnbaum, B. A., et al. “Extravasation detection accessory: clinical evaluation in 500 patients.Radiology, vol. 212, no. 2, Aug. 1999, pp. 431–38. Pubmed, doi:10.1148/radiology.212.2.r99au14431.
Birnbaum BA, Nelson RC, Chezmar JL, Glick SN. Extravasation detection accessory: clinical evaluation in 500 patients. Radiology. 1999 Aug;212(2):431–438.
Journal cover image

Published In

Radiology

DOI

ISSN

0033-8419

Publication Date

August 1999

Volume

212

Issue

2

Start / End Page

431 / 438

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Prospective Studies
  • Prevalence
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Injections, Intravenous
  • Humans
  • Female