Fast spin-echo imaging of the abdomen during breath-holding: an alternative to RASE and other fast imaging techniques.

Publication ,  Journal Article
Nelson, RC; Chezmar, JL
Published in: Gastrointest Radiol
1992

We have developed a rapidly acquired T1-weighted spin-echo pulse sequence that uses gradient echo-like parameters of TR 70 ms, TE 10 ms, NEX-1 (SE 70/10/1) with large pulse angles, and presaturation. This sequence yields two images of the abdomen during a comfortable breath-hold of 9 s. Preliminary phantom studies with this sequence demonstrated that peak signal-to-noise ratios occurred at pulse angles of 120 degrees and 135 degrees. Compared to this rapidly acquired sequence, a conventional T1-weighted spin-echo sequence of TR 140 ms, TE 10 ms, NEX-4 demonstrated 2.8 times the signal-to-noise ratio, 2.6 times the liver-spleen contrast-to-noise ratio but only 1.6 times the contrast-to-artifact ratio. When normalized for the imaging time, however, there was little difference in the signal-to-noise and contrast-to-noise ratios, although the SE 70/10/1 demonstrated 2.0 times the contrast-to-artifact ratio. We conclude that for abdominal imaging the SE 70/10/1 is an excellent alternative to RASE and other fast imaging techniques, and, although there are inherently low signal-to-noise ratios, it may be particularly useful when coupled with a paramagnetic contrast agent.

Duke Scholars

Published In

Gastrointest Radiol

DOI

ISSN

0364-2356

Publication Date

1992

Volume

17

Issue

4

Start / End Page

295 / 299

Location

United States

Related Subject Headings

  • Time Factors
  • Respiration
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Humans
  • Abdomen
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nelson, R. C., & Chezmar, J. L. (1992). Fast spin-echo imaging of the abdomen during breath-holding: an alternative to RASE and other fast imaging techniques. Gastrointest Radiol, 17(4), 295–299. https://doi.org/10.1007/BF01888573
Nelson, R. C., and J. L. Chezmar. “Fast spin-echo imaging of the abdomen during breath-holding: an alternative to RASE and other fast imaging techniques.Gastrointest Radiol 17, no. 4 (1992): 295–99. https://doi.org/10.1007/BF01888573.
Nelson, R. C., and J. L. Chezmar. “Fast spin-echo imaging of the abdomen during breath-holding: an alternative to RASE and other fast imaging techniques.Gastrointest Radiol, vol. 17, no. 4, 1992, pp. 295–99. Pubmed, doi:10.1007/BF01888573.

Published In

Gastrointest Radiol

DOI

ISSN

0364-2356

Publication Date

1992

Volume

17

Issue

4

Start / End Page

295 / 299

Location

United States

Related Subject Headings

  • Time Factors
  • Respiration
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Humans
  • Abdomen
  • 1103 Clinical Sciences