Skip to main content
Journal cover image

Recording characteristics of monopolar EMG electrodes.

Publication ,  Journal Article
Nandedkar, SD; Sanders, DB
Published in: Muscle Nerve
February 1991

Motor unit action potentials (MUAPs) and the electromyographic (EMG) interference pattern (IP) were recorded from the biceps muscle of 5 normal subjects using both a concentric needle (CN) and a disposable monopolar needle (MN) electrode. The MUAPs recorded by the MN electrode had higher amplitude and area and were more frequently complex than those recorded with the CN electrode. The MUAP duration and area: amplitude ratio were similar for both electrodes. Although the MN electrode had a larger recording surface, its dimensions (maximum diameter and length of the cone shaped tip) were similar to those of the CN electrode (minor and major axes of the elliptical recording tip). Based on these observations, we infer that the MN electrode may be more selective than the CN electrode, ie, the AP amplitude recorded by the MN electrode decreases faster than the AP amplitude recorded by a CN electrode when the distance of the muscle fiber from the recording electrode increases. Photomicrographs of the MN electrode after use demonstrated no evidence that the insulating material had peeled off. There was also no evidence that MUAP measurement values changed during the recordings as would be expected if the recording surface changed due to peeling of the insulating material.

Duke Scholars

Published In

Muscle Nerve

DOI

ISSN

0148-639X

Publication Date

February 1991

Volume

14

Issue

2

Start / End Page

108 / 112

Location

United States

Related Subject Headings

  • Surface Properties
  • Oscillometry
  • Neurology & Neurosurgery
  • Motor Neurons
  • Humans
  • Equipment Design
  • Electromyography
  • Electrodes, Implanted
  • Adult
  • Action Potentials
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nandedkar, S. D., & Sanders, D. B. (1991). Recording characteristics of monopolar EMG electrodes. Muscle Nerve, 14(2), 108–112. https://doi.org/10.1002/mus.880140204
Nandedkar, S. D., and D. B. Sanders. “Recording characteristics of monopolar EMG electrodes.Muscle Nerve 14, no. 2 (February 1991): 108–12. https://doi.org/10.1002/mus.880140204.
Nandedkar SD, Sanders DB. Recording characteristics of monopolar EMG electrodes. Muscle Nerve. 1991 Feb;14(2):108–12.
Nandedkar, S. D., and D. B. Sanders. “Recording characteristics of monopolar EMG electrodes.Muscle Nerve, vol. 14, no. 2, Feb. 1991, pp. 108–12. Pubmed, doi:10.1002/mus.880140204.
Nandedkar SD, Sanders DB. Recording characteristics of monopolar EMG electrodes. Muscle Nerve. 1991 Feb;14(2):108–112.
Journal cover image

Published In

Muscle Nerve

DOI

ISSN

0148-639X

Publication Date

February 1991

Volume

14

Issue

2

Start / End Page

108 / 112

Location

United States

Related Subject Headings

  • Surface Properties
  • Oscillometry
  • Neurology & Neurosurgery
  • Motor Neurons
  • Humans
  • Equipment Design
  • Electromyography
  • Electrodes, Implanted
  • Adult
  • Action Potentials