Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion.

Publication ,  Journal Article
Nelwan, SP; Kors, JA; Crater, SW; Meij, SH; van Dam, TB; Simoons, ML; Krucoff, MW
Published in: J Electrocardiol
2008

AIM: The aim of the study was to simultaneously test the EASI lead system and two other derived ECG methods against the standard 12-lead ECG during percutaneous coronary intervention (PCI). METHODS: During 44 percutaneous coronary interventions, a simultaneously recorded 12-lead and EASI ECG were marked at the start of the PCI (baseline) and at known ischemia caused by balloon inflation (peak). ST deviations were measured 60 ms after the J point at baseline and peak in all leads and were summated (SUMST) to assess overall changes. For regional changes, the lead with the highest ST deviation (PEAKST) was marked. For each patient, derived 12-lead ECGs were computed from the EASI leads and a lead subset using patient-specific coefficients (PS) and coefficients based on a patient population (GEN). Absolute differences were computed between each derived and routine ECG for SUMST and PEAKST. RESULTS: SUMST was at baseline 567 microV (range: 150-1707) and increased at peak to 871 microV (range: 350-2101). SUMST difference at peak was for EASI: 163 microV (CI: 90-236, P <.001), GEN: 46 microV (CI: 2-91, P = .40), and PS: 16 microV (CI: 3-30, P = .15). PEAKST difference at peak was for EASI: 49 microV (CI: 19-220, P = .02), GEN: 48 microV (CI: -43-154, P = .26), and PS: 20 microV (CI: -51-32, P = .65). CONCLUSION: Simultaneous direct comparison of three derived ECG methods shows overall and regional differences in accuracy across PS, GEN, and EASI. Median SUMST and PEAKST differences for PS are lower than for GEN and EASI, and show a more accurate reconstruction.

Duke Scholars

Published In

J Electrocardiol

DOI

EISSN

1532-8430

Publication Date

2008

Volume

41

Issue

3

Start / End Page

230 / 237

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Rest
  • Reproducibility of Results
  • Myocardial Ischemia
  • Humans
  • Electrodes
  • Electrocardiography
  • Coronary Artery Disease
  • Cardiovascular System & Hematology
  • Body Surface Potential Mapping
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nelwan, S. P., Kors, J. A., Crater, S. W., Meij, S. H., van Dam, T. B., Simoons, M. L., & Krucoff, M. W. (2008). Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion. J Electrocardiol, 41(3), 230–237. https://doi.org/10.1016/j.jelectrocard.2008.01.011
Nelwan, Stefan P., Jan A. Kors, Suzanne W. Crater, Simon H. Meij, Teus B. van Dam, Maarten L. Simoons, and Mitchell W. Krucoff. “Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion.J Electrocardiol 41, no. 3 (2008): 230–37. https://doi.org/10.1016/j.jelectrocard.2008.01.011.
Nelwan SP, Kors JA, Crater SW, Meij SH, van Dam TB, Simoons ML, et al. Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion. J Electrocardiol. 2008;41(3):230–7.
Nelwan, Stefan P., et al. “Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion.J Electrocardiol, vol. 41, no. 3, 2008, pp. 230–37. Pubmed, doi:10.1016/j.jelectrocard.2008.01.011.
Nelwan SP, Kors JA, Crater SW, Meij SH, van Dam TB, Simoons ML, Krucoff MW. Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion. J Electrocardiol. 2008;41(3):230–237.
Journal cover image

Published In

J Electrocardiol

DOI

EISSN

1532-8430

Publication Date

2008

Volume

41

Issue

3

Start / End Page

230 / 237

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Rest
  • Reproducibility of Results
  • Myocardial Ischemia
  • Humans
  • Electrodes
  • Electrocardiography
  • Coronary Artery Disease
  • Cardiovascular System & Hematology
  • Body Surface Potential Mapping