The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection

Publication ,  Journal Article
Nasab, NM; Samei, E
Published in: Progress in Biomedical Optics and Imaging - Proceedings of SPIE
June 30, 2006

In this paper, we evaluate the performance of biplane correlation imaging (BCI) using a set of off-angle projections acquired from an anthropomorphic chest phantom. BCI reduces the effect of anatomical noise, which would otherwise impact the detection subtle lesions in planar images. BCI also minimizes the number of false positives (FPs) when used in conjunction with computer aided diagnosis (CAD) applied to a set of coronal chest x-ray projections by eliminating non-correlated nodule candidates. In BCI, two digital images of the chest are acquired within a short time interval from two slightly different posterior projections. The image data are then incorporated into the CAD algorithm in which nodules are detected by examining the geometrical correlation of the detected signals in the two views, thus largely "canceling" the impact of anatomical noise. Seventy-one low exposure posterior projections were acquired of an anthropomorphic chest phantom containing tissue equivalent lesions with small angular separations (0.32 degree) over a range of 20 degrees, [-10°, +10°], along the vertical axis. The data were analyzed to determine the accuracy of the technique as a function of angular separation. The results indicated that the best performance was obtained when the angular separation of the projection pair was greater than 6 degrees. Within the range of optimum angular separation, the number of FPs per image, FPpI, was ∼1.1 with average sensitivity around 75% (supported by a grant from the NIH R01CA 109074).

Duke Scholars

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

Publication Date

June 30, 2006

Volume

6142 I
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nasab, N. M., & Samei, E. (2006). The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection. Progress in Biomedical Optics and Imaging - Proceedings of SPIE, 6142 I. https://doi.org/10.1117/12.652588
Nasab, N. M., and E. Samei. “The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection.” Progress in Biomedical Optics and Imaging - Proceedings of SPIE 6142 I (June 30, 2006). https://doi.org/10.1117/12.652588.
Nasab NM, Samei E. The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection. Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2006 Jun 30;6142 I.
Nasab, N. M., and E. Samei. “The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection.” Progress in Biomedical Optics and Imaging - Proceedings of SPIE, vol. 6142 I, June 2006. Scopus, doi:10.1117/12.652588.
Nasab NM, Samei E. The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection. Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2006 Jun 30;6142 I.

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

Publication Date

June 30, 2006

Volume

6142 I