3-D ultrasound guidance of surgical robotics: a feasibility study.

Publication ,  Journal Article
Pua, EC; Fronheiser, MP; Noble, JR; Light, ED; Wolf, PD; von Allmen, D; Smith, SW
Published in: IEEE transactions on ultrasonics, ferroelectrics, and frequency control
November 2006

Laparoscopic ultrasound has seen increased use as a surgical aide in general, gynecological, and urological procedures. The application of real-time, three-dimensional (RT3D) ultrasound to these laparoscopic procedures may increase information available to the surgeon and serve as an additional intraoperative guidance tool. The integration of RT3D with recent advances in robotic surgery also can increase automation and ease of use. In this study, a 1-cm diameter probe for RT3D has been used laparoscopically for in vivo imaging of a canine. The probe, which operates at 5 MHz, was used to image the spleen, liver, and gall bladder as well as to guide surgical instruments. Furthermore, the three-dimensional (3-D) measurement system of the volumetric scanner used with this probe was tested as a guidance mechanism for a robotic linear motion system in order to simulate the feasibility of RT3D/robotic surgery integration. Using images acquired with the 3-D laparoscopic ultrasound device, coordinates were acquired by the scanner and used to direct a robotically controlled needle toward desired in vitro targets as well as targets in a post-mortem canine. The rms error for these measurements was 1.34 mm using optical alignment and 0.76 mm using ultrasound alignment.

Duke Scholars

Published In

IEEE transactions on ultrasonics, ferroelectrics, and frequency control

DOI

EISSN

1525-8955

ISSN

0885-3010

Publication Date

November 2006

Volume

53

Issue

11

Start / End Page

1999 / 2008

Related Subject Headings

  • Ultrasonography, Interventional
  • Transducers
  • Systems Integration
  • Robotics
  • Laparoscopy
  • Laparoscopes
  • Imaging, Three-Dimensional
  • Feasibility Studies
  • Equipment Failure Analysis
  • Equipment Design
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Pua, E. C., Fronheiser, M. P., Noble, J. R., Light, E. D., Wolf, P. D., von Allmen, D., & Smith, S. W. (2006). 3-D ultrasound guidance of surgical robotics: a feasibility study. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 53(11), 1999–2008. https://doi.org/10.1109/tuffc.2006.140
Pua, Eric C., Matthew P. Fronheiser, Joanna R. Noble, Edward D. Light, Patrick D. Wolf, Daniel von Allmen, and Stephen W. Smith. “3-D ultrasound guidance of surgical robotics: a feasibility study.IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control 53, no. 11 (November 2006): 1999–2008. https://doi.org/10.1109/tuffc.2006.140.
Pua EC, Fronheiser MP, Noble JR, Light ED, Wolf PD, von Allmen D, et al. 3-D ultrasound guidance of surgical robotics: a feasibility study. IEEE transactions on ultrasonics, ferroelectrics, and frequency control. 2006 Nov;53(11):1999–2008.
Pua, Eric C., et al. “3-D ultrasound guidance of surgical robotics: a feasibility study.IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 53, no. 11, Nov. 2006, pp. 1999–2008. Epmc, doi:10.1109/tuffc.2006.140.
Pua EC, Fronheiser MP, Noble JR, Light ED, Wolf PD, von Allmen D, Smith SW. 3-D ultrasound guidance of surgical robotics: a feasibility study. IEEE transactions on ultrasonics, ferroelectrics, and frequency control. 2006 Nov;53(11):1999–2008.

Published In

IEEE transactions on ultrasonics, ferroelectrics, and frequency control

DOI

EISSN

1525-8955

ISSN

0885-3010

Publication Date

November 2006

Volume

53

Issue

11

Start / End Page

1999 / 2008

Related Subject Headings

  • Ultrasonography, Interventional
  • Transducers
  • Systems Integration
  • Robotics
  • Laparoscopy
  • Laparoscopes
  • Imaging, Three-Dimensional
  • Feasibility Studies
  • Equipment Failure Analysis
  • Equipment Design