WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector.

Publication ,  Journal Article
Belley, MD; Faught, A; Moore, B; Subashi, E; Langloss, B; Therien, MJ; Yoshizumi, TT; Chino, JP; Craciunescu, O
Published in: Med Phys
June 2016

PURPOSE: Development of a novel on-line dosimetry tool is needed to move toward patient-specific quality assurance measurements for Ir-192 HDR brachytherapy to verify accurate dose delivery to the intended location. This work describes the development and use of a nano-crystalline yttrium oxide inorganic scintillator based optical-fiber detector capable of acquiring real-time high-precision dose measurements during tandem and ovoid (T&O) gynecological (GYN) applicator Ir-192 HDR brachytherapy procedures. METHODS: An optical-fiber detector was calibrated by acquiring light output measurements in liquid water at 3, 5, 7, and 9cm radial source-detector-distances from an Ir-192 HDR source. A regression model was fit to the data to describe the relative light output per unit dose (TG-43 derived) as a function of source-detector-distance. Next, the optical-fiber detector was attached to a vaginal balloon fixed to a Varian Fletcher-Suit-Delclos-style applicator (to mimic clinical setup), and localized by acquiring high-resolution computed tomography (CT) images. To compare the physical point dose to the TPS calculated values (TG-43 and Acuros-BV), a phantom measurement was performed, by submerging the T&O applicator in a liquid water bath and delivering a treatment template representative of a clinical T&O procedure. The fiber detector collected scintillation signal as a function of time, and the calibration data was applied to calculate both real-time dose rate, and cumulative dose. RESULTS: Fiber cumulative dose values were 100.0cGy, 94.3cGy, and 348.9cGy from the tandem, left ovoid, and right ovoid dwells, respectively (total of 443.2cGy). A plot of real time dose rate during the treatment was also acquired. The TPS values at the fiber location were 458.4cGy using TG-43, and 437.6cGy using Acuros-BV calculated as Dm,m (per TG-186). CONCLUSION: The fiber measured dose value agreement was 3% vs TG-43 and -1% vs Acuros-BV. This fiber detector opens up new possibilities for performing patient-specific quality assurance for Ir-192 HDR GYN procedures. Funding from Coulter Foundation, Duke Bio-medical Engineering. Company is being created around the detector technology. Duke holds patents on the technology.

Duke Scholars

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

June 2016

Volume

43

Issue

6

Start / End Page

3809 / 3810

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Belley, M. D., Faught, A., Moore, B., Subashi, E., Langloss, B., Therien, M. J., … Craciunescu, O. (2016). WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector. Med Phys, 43(6), 3809–3810. https://doi.org/10.1118/1.4957812
Belley, M. D., A. Faught, B. Moore, E. Subashi, B. Langloss, M. J. Therien, T. T. Yoshizumi, J. P. Chino, and O. Craciunescu. “WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector.Med Phys 43, no. 6 (June 2016): 3809–10. https://doi.org/10.1118/1.4957812.
Belley MD, Faught A, Moore B, Subashi E, Langloss B, Therien MJ, et al. WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector. Med Phys. 2016 Jun;43(6):3809–10.
Belley, M. D., et al. “WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector.Med Phys, vol. 43, no. 6, June 2016, pp. 3809–10. Pubmed, doi:10.1118/1.4957812.
Belley MD, Faught A, Moore B, Subashi E, Langloss B, Therien MJ, Yoshizumi TT, Chino JP, Craciunescu O. WE-DE-201-07: Measurement of Real-Time Dose for Tandem and Ovoid Brachytherapy Procedures Using a High Precision Optical Fiber Radiation Detector. Med Phys. 2016 Jun;43(6):3809–3810.

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

June 2016

Volume

43

Issue

6

Start / End Page

3809 / 3810

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences