Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control.

Publication ,  Journal Article
El-Chami, MF; Al-Samadi, F; Clementy, N; Garweg, C; Martinez-Sande, JL; Piccini, JP; Iacopino, S; Lloyd, M; Viñolas Prat, X; Jacobsen, MD ...
Published in: Heart Rhythm
December 2018

BACKGROUND: Early results of the Micra Investigational Device Exemption (IDE) study and Micra Post-Approval Registry (PAR) demonstrated excellent safety and efficacy performance; however, intermediate-term results across a large patient population in the real-world setting have not been evaluated. OBJECTIVES: We report updated performance of the Micra transcatheter pacemaker from a worldwide PAR and compare it with the IDE study as well as a transvenous historical control. METHODS: The safety objective of the analysis was system- or procedure-related major complications through 12 months postimplantation. We compared the major complication rate with that of the 726 patients from the IDE and with a reference data set of 2667 patients with transvenous pacemakers by using a Fine-Gray competing risk model. RESULTS: The Micra device was successfully implanted in 1801 of 1817 patients (99.1%). The mean follow-up period was 6.8 ± 6.9 months. Through 12 months, the major complication rate was 2.7% (95% confidence interval [CI] 2.0%-3.7%). The risk of major complications for Micra PAR patients was 63% lower than that for patients with transvenous pacemakers through 12 months postimplantation (hazard ratio 0.37; 95% CI 0.27-0.52; P < .001). The major complication rate trended lower in the PAR than in the IDE study (hazard ratio 0.71; 95% CI 0.44-1.1; P = .160), driven by the lower pericardial effusion rate in the PAR. There were 3 cases of infection associated with the procedure, but none required device removal and there were no battery or telemetry issues. Pacing thresholds were low and stable through 12 months postimplantation. CONCLUSION: Performance of the Micra transcatheter pacemaker in international clinical practice remains consistent with previously reported data. Major complications were infrequent and occurred 63% less often compared to transvenous systems. CLINICAL TRIAL REGISTRATION: Micra Transcatheter Pacing System Post-Approval Registry ClinicalTrials.gov identifier: NCT02536118; Micra Transcatheter Pacing Study ClinicalTrials.gov identifier: NCT02004873.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

December 2018

Volume

15

Issue

12

Start / End Page

1800 / 1807

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Time Factors
  • Registries
  • Prospective Studies
  • Pacemaker, Artificial
  • Male
  • Humans
  • Follow-Up Studies
  • Female
  • Equipment Design
 

Citation

APA
Chicago
ICMJE
MLA
NLM
El-Chami, M. F., Al-Samadi, F., Clementy, N., Garweg, C., Martinez-Sande, J. L., Piccini, J. P., … Roberts, P. R. (2018). Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control. Heart Rhythm, 15(12), 1800–1807. https://doi.org/10.1016/j.hrthm.2018.08.005
El-Chami, Mikhael F., Faisal Al-Samadi, Nicolas Clementy, Christophe Garweg, Jose Luis Martinez-Sande, Jonathan P. Piccini, Saverio Iacopino, et al. “Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control.Heart Rhythm 15, no. 12 (December 2018): 1800–1807. https://doi.org/10.1016/j.hrthm.2018.08.005.
El-Chami MF, Al-Samadi F, Clementy N, Garweg C, Martinez-Sande JL, Piccini JP, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018 Dec;15(12):1800–7.
El-Chami, Mikhael F., et al. “Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control.Heart Rhythm, vol. 15, no. 12, Dec. 2018, pp. 1800–07. Pubmed, doi:10.1016/j.hrthm.2018.08.005.
El-Chami MF, Al-Samadi F, Clementy N, Garweg C, Martinez-Sande JL, Piccini JP, Iacopino S, Lloyd M, Viñolas Prat X, Jacobsen MD, Ritter P, Johansen JB, Tondo C, Liu F, Fagan DH, Eakley AK, Roberts PR. Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018 Dec;15(12):1800–1807.
Journal cover image

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

December 2018

Volume

15

Issue

12

Start / End Page

1800 / 1807

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Time Factors
  • Registries
  • Prospective Studies
  • Pacemaker, Artificial
  • Male
  • Humans
  • Follow-Up Studies
  • Female
  • Equipment Design