Measurement of methotrexate polyglutamates in human erythrocytes by ion-pair UPLC-MS/MS.
BACKGROUND: Low-dose methotrexate is used for the treatment of rheumatoid arthritis and juvenile idiopathic arthritis, but its effectiveness greatly varies between individuals. Therapeutic drug monitoring of intracellular methotrexate metabolites, the γ-polyglutamates (MTXGlu(n)), in human erythrocytes has shown promise in providing a basis for individualization of therapy. RESULTS: This work presents expedient methodology for the analysis of MTXGlu(1-7) in human erythrocytes by ion-pair UPLC with detection by tandem MS (UPLC-ESI-MS/MS). The use of N,N-dimethylheptylamine as an ion-pair agent was found to be favorable over others. Thermal extraction of erythrocyte lysates provides a simple one-step extraction procedure. The entire chromatographic run time is 6 min and the assay was validated within the therapeutic range of these metabolites CONCLUSION: The developed sample preparation procedure in combination with ion-pair UPLC-ESI-MS/MS analysis allowed for expedient quantitation of MTXGlu(1-7) in human erythrocytes. The rapid analysis time would enable therapeutic drug monitoring of MTXGlu(1-7) in the clinic.
Duke Scholars
Published In
DOI
EISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Tandem Mass Spectrometry
- Reproducibility of Results
- Polyglutamic Acid
- Phenethylamines
- Methotrexate
- Isotope Labeling
- Ions
- Humans
- Erythrocytes
- Deuterium
Citation

Published In
DOI
EISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Tandem Mass Spectrometry
- Reproducibility of Results
- Polyglutamic Acid
- Phenethylamines
- Methotrexate
- Isotope Labeling
- Ions
- Humans
- Erythrocytes
- Deuterium