Skip to main content

Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry.

Publication ,  Journal Article
Sandhu, A; Varosy, PD; Du, C; Aleong, RG; Tumolo, AZ; West, JJ; Tzou, WS; Curtis, JP; Freeman, JV; Friedman, DJ; Hess, PL
Published in: Circ Cardiovasc Interv
December 2022

BACKGROUND: Left atrial appendage occlusion is an important alternative to anticoagulation in select patients with nonvalvular atrial fibrillation. Trends in real-world device sizing and associated short-term complications have not been characterized. METHODS: Using the National Cardiovascular Data Left Atrial Appendage Occlusion (NCDR LAAO) Registry, patients who underwent left atrial appendage occlusion with a Watchman 2.5 device from January 1, 2016, to June 30, 2020, were identified. Patients were stratified by device size based on left atrial appendage orifice size, and categorized as receiving a device that was undersized, oversized, or per manufacturer recommendation. Relationships between device sizing and short-term outcomes, including pericardial effusion, device embolism, and significant leak, were assessed. RESULTS: Of the 68 456 patients, 6539 (10.5%) of patients received undersized devices, 17 791 (26.0%) according to manufacturer recommendations, and 44 126 (64.4%) received an oversized device. The 27-mm device was most commonly deployed [21 736 (31.8%)], whereas the smallest and largest devices (21 and 33 mm) were least commonly deployed [7695 (11.2%) and 9077 (13.3%), respectively]. Compared with manufacturer recommended sizing, there was no difference in the odds of pericardial effusion for either undersized (1.048 [95% CI' 0.801-1.372]; P=0.733) or oversized (1.101 [95% CI' 0.933-1.298]; P=0.254) devices. Similarly, relative to manufacturer recommended sizing, the odds of a composite adverse outcome of device migration or embolization and significant peridevice leak at 45 days were similar among undersized devices (1.030 [95% CI' 0.735-1.444]; P=0.863) and favorable for oversized devices (0.701 [95% CI' 0.561-0.876]; P=0.002) devices, primarily driven by lower odds of leak. Selection of oversized devices increased significantly over the study period (from 60.3% in 2016 to 66.0% in 2020; P<0.001). CONCLUSIONS: Among patients undergoing left atrial appendage occlusion with the first-generation Watchman device, receipt of oversized devices was common and increased over time. The high prevalence of oversizing was associated with lower odds of significant leak or device embolization without increased odds of other adverse events.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Circ Cardiovasc Interv

DOI

EISSN

1941-7632

Publication Date

December 2022

Volume

15

Issue

12

Start / End Page

e012183

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Stroke
  • Registries
  • Pericardial Effusion
  • Humans
  • Cardiovascular System & Hematology
  • Cardiac Catheterization
  • Atrial Fibrillation
  • Atrial Appendage
  • 3201 Cardiovascular medicine and haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Sandhu, A., Varosy, P. D., Du, C., Aleong, R. G., Tumolo, A. Z., West, J. J., … Hess, P. L. (2022). Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry. Circ Cardiovasc Interv, 15(12), e012183. https://doi.org/10.1161/CIRCINTERVENTIONS.122.012183
Sandhu, Amneet, Paul D. Varosy, Chengan Du, Ryan G. Aleong, Alexis Z. Tumolo, J Jason West, Wendy S. Tzou, et al. “Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry.Circ Cardiovasc Interv 15, no. 12 (December 2022): e012183. https://doi.org/10.1161/CIRCINTERVENTIONS.122.012183.
Sandhu A, Varosy PD, Du C, Aleong RG, Tumolo AZ, West JJ, et al. Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry. Circ Cardiovasc Interv. 2022 Dec;15(12):e012183.
Sandhu, Amneet, et al. “Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry.Circ Cardiovasc Interv, vol. 15, no. 12, Dec. 2022, p. e012183. Pubmed, doi:10.1161/CIRCINTERVENTIONS.122.012183.
Sandhu A, Varosy PD, Du C, Aleong RG, Tumolo AZ, West JJ, Tzou WS, Curtis JP, Freeman JV, Friedman DJ, Hess PL. Device-Sizing and Associated Complications With Left Atrial Appendage Occlusion: Findings From the NCDR LAAO Registry. Circ Cardiovasc Interv. 2022 Dec;15(12):e012183.

Published In

Circ Cardiovasc Interv

DOI

EISSN

1941-7632

Publication Date

December 2022

Volume

15

Issue

12

Start / End Page

e012183

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Stroke
  • Registries
  • Pericardial Effusion
  • Humans
  • Cardiovascular System & Hematology
  • Cardiac Catheterization
  • Atrial Fibrillation
  • Atrial Appendage
  • 3201 Cardiovascular medicine and haematology