Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements.

Publication ,  Conference
Lundstrom, JA; Khemani, RG; Hotz, J; Newth, CJL; Achanta, S; Gentile, MA; Hedin, DS
Published in: Annu Int Conf IEEE Eng Med Biol Soc
July 2022

There is growing research showing the importance of measuring esophageal pressure as a surrogate for pleural pressure for patients on mechanical ventilators. The most common measurement method uses a balloon catheter, whose accuracy can vary based on patient anatomy, balloon position, balloon inflation, and the presence of other tubes in the esophagus. The authors present the development and initial testing results of a new combination catheter, utilizing fiberoptic pressure sensing to provide more accurate esophageal pressure measurements and allowing for the incorporation of a feeding tube and temperature sensor.

Duke Scholars

Published In

Annu Int Conf IEEE Eng Med Biol Soc

DOI

EISSN

2694-0604

Publication Date

July 2022

Volume

2022

Start / End Page

3269 / 3272

Location

United States

Related Subject Headings

  • Pressure
  • Humans
  • Fiber Optic Technology
  • Esophagus
  • Catheters
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Lundstrom, J. A., Khemani, R. G., Hotz, J., Newth, C. J. L., Achanta, S., Gentile, M. A., & Hedin, D. S. (2022). Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements. In Annu Int Conf IEEE Eng Med Biol Soc (Vol. 2022, pp. 3269–3272). United States. https://doi.org/10.1109/EMBC48229.2022.9871790
Lundstrom, Julie A., Robinder G. Khemani, Justin Hotz, Christopher J. L. Newth, Satyanarayana Achanta, Michael A. Gentile, and Daniel S. Hedin. “Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements.” In Annu Int Conf IEEE Eng Med Biol Soc, 2022:3269–72, 2022. https://doi.org/10.1109/EMBC48229.2022.9871790.
Lundstrom JA, Khemani RG, Hotz J, Newth CJL, Achanta S, Gentile MA, et al. Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements. In: Annu Int Conf IEEE Eng Med Biol Soc. 2022. p. 3269–72.
Lundstrom, Julie A., et al. “Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements.Annu Int Conf IEEE Eng Med Biol Soc, vol. 2022, 2022, pp. 3269–72. Pubmed, doi:10.1109/EMBC48229.2022.9871790.
Lundstrom JA, Khemani RG, Hotz J, Newth CJL, Achanta S, Gentile MA, Hedin DS. Development of a Pleural Pressure Catheter via Continuous Fiberoptic Esophageal Pressure Measurements. Annu Int Conf IEEE Eng Med Biol Soc. 2022. p. 3269–3272.

Published In

Annu Int Conf IEEE Eng Med Biol Soc

DOI

EISSN

2694-0604

Publication Date

July 2022

Volume

2022

Start / End Page

3269 / 3272

Location

United States

Related Subject Headings

  • Pressure
  • Humans
  • Fiber Optic Technology
  • Esophagus
  • Catheters