Skip to main content
Journal cover image

Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming.

Publication ,  Journal Article
Overson, DK; Darnell, D; Robb, F; Song, AW; Truong, T-K
Published in: Magn Reson Med
February 2024

PURPOSE: To develop a flexible, lightweight, and multi-purpose integrated parallel reception, excitation, and shimming (iPRES) coil array that can conform to the subject's anatomy and perform MR imaging and localized B0 shimming in different anatomical regions with a high SNR, shimming performance, ease of positioning, and subject comfort. METHODS: A four-channel flexible iPRES coil array was constructed by enabling RF and direct currents to flow on the same flexible coil elements for imaging and shimming, respectively. Shimming experiments were performed with the coil array wrapped around the knee or neck of healthy subjects to demonstrate its high shimming performance and versatility. Additionally, its SNR and shimming performance in the knee were compared to those obtained with the coil array wrapped around a larger rigid tube designed to fit most knee sizes. RESULTS: Shimming with the coil array wrapped around the knee or neck resulted in an average reduction in B0 RMSE of 50.1% and 40.5% relative to first-order and second-order spherical harmonic shimming, respectively, and substantially reduced distortions in DWI images. In contrast, shimming the knee with the coil array wrapped around the rigid tube only provided a 29.6% reduction in B0 RMSE, whereas the SNR was reduced by 58.7%. CONCLUSION: The flexible iPRES coil array can conform to different anatomical regions and perform imaging and localized B0 shimming with a higher SNR, shimming performance, ease of positioning, and comfort compared to a rigid iPRES coil array, which should be valuable for many applications throughout the human body.

Duke Scholars

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

February 2024

Volume

91

Issue

2

Start / End Page

842 / 849

Location

United States

Related Subject Headings

  • Radio Waves
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Knee Joint
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Overson, D. K., Darnell, D., Robb, F., Song, A. W., & Truong, T.-K. (2024). Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming. Magn Reson Med, 91(2), 842–849. https://doi.org/10.1002/mrm.29891
Overson, Devon Karl, Dean Darnell, Fraser Robb, Allen W. Song, and Trong-Kha Truong. “Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming.Magn Reson Med 91, no. 2 (February 2024): 842–49. https://doi.org/10.1002/mrm.29891.
Overson DK, Darnell D, Robb F, Song AW, Truong T-K. Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming. Magn Reson Med. 2024 Feb;91(2):842–9.
Overson, Devon Karl, et al. “Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming.Magn Reson Med, vol. 91, no. 2, Feb. 2024, pp. 842–49. Pubmed, doi:10.1002/mrm.29891.
Overson DK, Darnell D, Robb F, Song AW, Truong T-K. Flexible multi-purpose integrated RF/shim coil array for MRI and localized B0 shimming. Magn Reson Med. 2024 Feb;91(2):842–849.
Journal cover image

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

February 2024

Volume

91

Issue

2

Start / End Page

842 / 849

Location

United States

Related Subject Headings

  • Radio Waves
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Knee Joint
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • 4003 Biomedical engineering
  • 0903 Biomedical Engineering