Skip to main content

Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST)

Publication ,  Conference
Song, P; Macdonald, MC; Behler, RH; Lanning, JD; Wang, MH; Urban, MW; Manduca, A; Zhao, H; Callstrom, MR; Alizad, A; Greenleaf, JF; Chen, S
Published in: IEEE International Ultrasonics Symposium, IUS
October 20, 2014

Two-dimensional (2D) shear wave elastography has emerged as a popular clinical and research tool for a variety of applications. Implementing 2D shear wave elastography on a conventional ultrasound scanner, however, is a challenging task due to the low imaging frame rate, which is essentially limited by the low parallel beamforming capability provided by the hardware beamformers on these scanners. This study developed a time aligned sequential tracking (TAST) technique that enables high pulse repetition frequency (PRF) shear wave tracking on conventional ultrasound scanners. The Comb-push Ultrasound Shear Elastography (CUSE) technique was combined with TAST to realize large field-of-view (FOV) 2D shear wave elastography. CUSE and TAST were implemented on the General Electric LOGIQ E9 (LE9) scanner. Phantom experiments showed that the LE9 could reconstruct large FOV, high quality 2D shear wave speed maps with accurate shear wave speed measurements that are in good agreement to other 2D shear wave imaging techniques. An inclusion phantom study showed that LE9 had comparable performance to the Aixplorer (Supersonic Imagine) in terms of bias and precision in measuring different sized inclusions. Finally, in vivo case studies of a breast with a malignant mass, and a liver from a healthy subject demonstrated the feasibility of using LE9 for in vivo 2D shear wave elastography.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

IEEE International Ultrasonics Symposium, IUS

DOI

EISSN

1948-5727

ISSN

1948-5719

Publication Date

October 20, 2014

Start / End Page

1101 / 1104
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Song, P., Macdonald, M. C., Behler, R. H., Lanning, J. D., Wang, M. H., Urban, M. W., … Chen, S. (2014). Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST). In IEEE International Ultrasonics Symposium, IUS (pp. 1101–1104). https://doi.org/10.1109/ULTSYM.2014.0270
Song, P., M. C. Macdonald, R. H. Behler, J. D. Lanning, M. H. Wang, M. W. Urban, A. Manduca, et al. “Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST).” In IEEE International Ultrasonics Symposium, IUS, 1101–4, 2014. https://doi.org/10.1109/ULTSYM.2014.0270.
Song P, Macdonald MC, Behler RH, Lanning JD, Wang MH, Urban MW, et al. Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST). In: IEEE International Ultrasonics Symposium, IUS. 2014. p. 1101–4.
Song, P., et al. “Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST).” IEEE International Ultrasonics Symposium, IUS, 2014, pp. 1101–04. Scopus, doi:10.1109/ULTSYM.2014.0270.
Song P, Macdonald MC, Behler RH, Lanning JD, Wang MH, Urban MW, Manduca A, Zhao H, Callstrom MR, Alizad A, Greenleaf JF, Chen S. Shear wave elastography on the GE LOGIQ E9 with Comb-push Ultrasound Shear Elastography (CUSE) and time aligned sequential tracking (TAST). IEEE International Ultrasonics Symposium, IUS. 2014. p. 1101–1104.

Published In

IEEE International Ultrasonics Symposium, IUS

DOI

EISSN

1948-5727

ISSN

1948-5719

Publication Date

October 20, 2014

Start / End Page

1101 / 1104