Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA.

Publication ,  Journal Article
Friedman, DJ; Shadrin, I; Goldbarg, S; Trulock, KM; Patel, A; Loring, Z; Coles, SA; Gandhi, G; Upadhyay, GA; Wold, N; Jones, PW; Ruble, SB ...
Published in: Heart Rhythm
August 2025

BACKGROUND: Left bundle branch area pacing (LBBAP) has been rapidly adopted despite a lack of leads with approved indications for LBBAP. OBJECTIVE: This study aims to analyze the performance of a 6Fr active fixation, stylet-driven lead in LBBAP. METHODS: The multicenter INSIGHT-LBBA registry analyzed consecutive LBBAP implant attempts with INGEVITY+ leads. Data were retrospectively collected with standardized entry via REDCap. Prespecified safety (lead-related complication [LRC]-free rate with performance goal >90% at 3 months) and effectiveness (pacing capture threshold [PCT] ≤2 V@ 0.4 ms and R-waves ≥5 mV; 3-month performance goal = 80%) endpoints were analyzed. External validation of lead effectiveness was performed on a separate cohort of LBBAP patients from the LATITUDE remote monitoring database. RESULTS: The study enrolled 1122 single- or dual-chamber pacemaker patients from 8 US sites (mean age 76 ±10 years, 43% female, 48% atrioventicular [AV] block, 40% sinus node dysfunction). Those successfully implanted had a median follow-up of 302 days. Left bundle or left septal capture was achieved in 95.6% of patients (1073 of 1122) and 89.6% of leads (1073 of 1197, 1.07 leads per patient). The 3-month LRC-free rate was 97.7% (lower confidence limit [LCL] 96.6%) with an LRC-free rate of 97.2% at 24 months. At 3 months, 98.8% (LCL 97.7%) of PCTs were ≤2 V, and 94.8% (LCL 92.7%) of sensed amplitudes were ≥ 5 mV. Results were stable through 24 months; 864 LBBAP implants were analyzed in the external validation cohort, and lead effectiveness was clinically indistinguishable across datasets. CONCLUSION: In this study of an active fixation, stylet-driven lead, all prespecified performance goals for pacing, sensing, and LRC rates were met when used for LBBAP. External validation demonstrated concordant results. CLINICAL TRIAL REGISTRATION: NCT06014866.

Duke Scholars

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

August 2025

Volume

22

Issue

8

Start / End Page

2047 / 2054

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Retrospective Studies
  • Registries
  • Pacemaker, Artificial
  • Male
  • Humans
  • Follow-Up Studies
  • Female
  • Equipment Design
  • Electrocardiography
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Friedman, D. J., Shadrin, I., Goldbarg, S., Trulock, K. M., Patel, A., Loring, Z., … Latanich, C. A. (2025). Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA. Heart Rhythm, 22(8), 2047–2054. https://doi.org/10.1016/j.hrthm.2025.01.041
Friedman, Daniel J., Ilya Shadrin, Seth Goldbarg, Kevin M. Trulock, Apoor Patel, Zak Loring, Sara A. Coles, et al. “Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA.Heart Rhythm 22, no. 8 (August 2025): 2047–54. https://doi.org/10.1016/j.hrthm.2025.01.041.
Friedman DJ, Shadrin I, Goldbarg S, Trulock KM, Patel A, Loring Z, et al. Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA. Heart Rhythm. 2025 Aug;22(8):2047–54.
Friedman, Daniel J., et al. “Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA.Heart Rhythm, vol. 22, no. 8, Aug. 2025, pp. 2047–54. Pubmed, doi:10.1016/j.hrthm.2025.01.041.
Friedman DJ, Shadrin I, Goldbarg S, Trulock KM, Patel A, Loring Z, Coles SA, Gandhi G, Upadhyay GA, Wold N, Jones PW, Ruble SB, Weinstock J, Latanich CA. Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBA. Heart Rhythm. 2025 Aug;22(8):2047–2054.
Journal cover image

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

August 2025

Volume

22

Issue

8

Start / End Page

2047 / 2054

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Retrospective Studies
  • Registries
  • Pacemaker, Artificial
  • Male
  • Humans
  • Follow-Up Studies
  • Female
  • Equipment Design
  • Electrocardiography