Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure.

Publication ,  Journal Article
Shah, MJ; Nehgme, R; Carboni, M; Murphy, JD
Published in: Pacing Clin Electrophysiol
July 2004

The purpose of the study was to investigate the results of endocardial lead implantation, lead performance, and follow-up in young patients after the Fontan procedure. A retrospective study was conducted with patients who had endocardial atrial pacing for SND and intact AVN function after Fontan from two pediatric centers. Patient demographics, pacing, and sensing data of endocardial atrial leads were analyzed at the time of pacemaker implantation and follow-up visits. Fifteen patients (weight 42.6 +/- 35 kg) had transvenous endocardial atrial lead implantation at an average age of 11.4 +/- 6.5 years. Active-fixation leads were used in all patients and steroid elution was present in 12 (80%) patients. Adequate P wave sensing was obtained in patients with sinus rhythm (n = 10); the remaining four patients had junctional rhythm without measurable P waves. Lead failure was not observed in any patient during the follow-up period of 2.9 +/- 2.1 years. The energy threshold at implantation was 1.46 +/- 1.5 microJ, 1.54 +/- 0.75 microJ at 3 months, 0.62 +/- 0.45 microJ at 1 year, 0.72 +/- 0.65 microJ at 2 years, 0.75 +/- 0.55 microJ at 3 years, and 0.8 +/- 0.85 microJ at 5 years postimplant. The lead impedance was 648 +/- 298 omega at implantation, 714 +/- 163 omega at 3 months, 744 +/- 195 omega at 1 year, 734 +/- 198 omega at 2 years, 800 +/- 142 omega at 3 years and 830 +/- 200 omega 5 years postimplant. Anticoagulation therapy (aspirin n = 5, warfarin n = 8) was continued by 13 patients. Complications consisted of a pneumothorax at implantation and a transient ischemic attack in one patient 4 years after pacemaker implant. Endocardial atrial leads offer low energy thresholds and can be implanted relatively safely in Fontan patients.

Duke Scholars

Published In

Pacing Clin Electrophysiol

DOI

ISSN

0147-8389

Publication Date

July 2004

Volume

27

Issue

7

Start / End Page

949 / 954

Location

United States

Related Subject Headings

  • Pacemaker, Artificial
  • Infant
  • Humans
  • Heart Atria
  • Fontan Procedure
  • Electrodes, Implanted
  • Child, Preschool
  • Child
  • Cardiovascular System & Hematology
  • Cardiac Pacing, Artificial
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Shah, M. J., Nehgme, R., Carboni, M., & Murphy, J. D. (2004). Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure. Pacing Clin Electrophysiol, 27(7), 949–954. https://doi.org/10.1111/j.1540-8159.2004.00564.x
Shah, Maully J., Rodrigo Nehgme, Michael Carboni, and John D. Murphy. “Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure.Pacing Clin Electrophysiol 27, no. 7 (July 2004): 949–54. https://doi.org/10.1111/j.1540-8159.2004.00564.x.
Shah MJ, Nehgme R, Carboni M, Murphy JD. Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure. Pacing Clin Electrophysiol. 2004 Jul;27(7):949–54.
Shah, Maully J., et al. “Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure.Pacing Clin Electrophysiol, vol. 27, no. 7, July 2004, pp. 949–54. Pubmed, doi:10.1111/j.1540-8159.2004.00564.x.
Shah MJ, Nehgme R, Carboni M, Murphy JD. Endocardial atrial pacing lead implantation and midterm follow-up in young patients with sinus node dysfunction after the fontan procedure. Pacing Clin Electrophysiol. 2004 Jul;27(7):949–954.
Journal cover image

Published In

Pacing Clin Electrophysiol

DOI

ISSN

0147-8389

Publication Date

July 2004

Volume

27

Issue

7

Start / End Page

949 / 954

Location

United States

Related Subject Headings

  • Pacemaker, Artificial
  • Infant
  • Humans
  • Heart Atria
  • Fontan Procedure
  • Electrodes, Implanted
  • Child, Preschool
  • Child
  • Cardiovascular System & Hematology
  • Cardiac Pacing, Artificial