A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation.

Publication ,  Journal Article
Jain, MK; Wolf, PD
Published in: Annals of biomedical engineering
September 2000

A novel three-dimensional finite element model for the study of radiofrequency ablation is presented. The model was used to perform an analysis of the temperature distribution in a tissue block heated by RF energy and cooled by blood (fluid) flow. This work extends earlier models by including true flow in place of a convective boundary condition to simulate realistic experimental conditions and to improve the prediction of blood temperatures. The effect of fluid flow on the temperature distribution, the lesion dimensions, and the ablation efficiency was studied. Three flow velocities were simulated: (i) 30, (ii) 55, and (iii) 85 mm/s. The modeling results were validated qualitatively and quantitatively with in vitro data. The correlation coefficients between the modeling and the experimental temperature measurements were 0.98, 0.97, and 0.95 for flows (i)-(iii), respectively. The slopes were 0.89, 0.95, and 1.06, and the mean root mean square differences between modeling and experimental temperature measurements were 17.3% +/- 11.6%, 15.8% +/- 13.4%, and 18.8% +/- 14.9% for flows (i)-(iii), respectively. A comparison of temperature distribution obtained with a convective boundary versus inclusion of fluid motion showed that the convective boundary resulted in a similar tissue temperature distribution, but overestimated fluid temperatures and lacked the flow asymmetry seen in the true flow model.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Annals of biomedical engineering

DOI

EISSN

1573-9686

ISSN

0090-6964

Publication Date

September 2000

Volume

28

Issue

9

Start / End Page

1075 / 1084

Related Subject Headings

  • Temperature
  • Radiofrequency Therapy
  • Models, Cardiovascular
  • Humans
  • Computer Simulation
  • Catheter Ablation
  • Blood Flow Velocity
  • Biomedical Engineering
  • Biomedical Engineering
  • Arrhythmias, Cardiac
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Jain, M. K., & Wolf, P. D. (2000). A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation. Annals of Biomedical Engineering, 28(9), 1075–1084. https://doi.org/10.1114/1.1310219
Jain, M. K., and P. D. Wolf. “A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation.Annals of Biomedical Engineering 28, no. 9 (September 2000): 1075–84. https://doi.org/10.1114/1.1310219.
Jain MK, Wolf PD. A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation. Annals of biomedical engineering. 2000 Sep;28(9):1075–84.
Jain, M. K., and P. D. Wolf. “A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation.Annals of Biomedical Engineering, vol. 28, no. 9, Sept. 2000, pp. 1075–84. Epmc, doi:10.1114/1.1310219.
Jain MK, Wolf PD. A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation. Annals of biomedical engineering. 2000 Sep;28(9):1075–1084.
Journal cover image

Published In

Annals of biomedical engineering

DOI

EISSN

1573-9686

ISSN

0090-6964

Publication Date

September 2000

Volume

28

Issue

9

Start / End Page

1075 / 1084

Related Subject Headings

  • Temperature
  • Radiofrequency Therapy
  • Models, Cardiovascular
  • Humans
  • Computer Simulation
  • Catheter Ablation
  • Blood Flow Velocity
  • Biomedical Engineering
  • Biomedical Engineering
  • Arrhythmias, Cardiac