Variable compensation chest radiography performed with a computed radiography system: design considerations and initial clinical experience.
The authors describe a variable compensation (VC) technique in which an x-ray equalizer and a computed radiography system are used. The VC technique allows retrospective alteration of equalized chest appearance with maintenance of improved signal-to-noise ratio in dense regions. Two imaging plates are used: one upstream of the patient to record the incident beam profile and one down-stream to record the equalized image. Subtraction of a weighted version of the upstream image from the down-stream image permits alteration of the appearance of the VC image, from the extremes of stimulated-unequalized to highly equalized. VC image appearance was optimized with a real-time workstation. The quality of VC images obtained in 33 patients was evaluated by three chest radiologists. Mediastinal appearance was better on VC equalized images than on conventional screen-film images. The stimulation of the appearance of a conventional radiograph with VC proved useful in interpretation of lung appearances on equalized radiographs.
Duke Scholars
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Tomography, X-Ray Computed
- Technology, Radiologic
- Radiography, Thoracic
- Radiographic Image Enhancement
- Nuclear Medicine & Medical Imaging
- Middle Aged
- Male
- Humans
- Aged
- 3202 Clinical sciences
Citation
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Tomography, X-Ray Computed
- Technology, Radiologic
- Radiography, Thoracic
- Radiographic Image Enhancement
- Nuclear Medicine & Medical Imaging
- Middle Aged
- Male
- Humans
- Aged
- 3202 Clinical sciences