Skip to main content
Journal cover image

Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities.

Publication ,  Journal Article
Fahey, BJ; Hsu, SJ; Wolf, PD; Nelson, RC; Trahey, GE
Published in: Phys Med Biol
August 7, 2006

Previous studies have established the feasibility of monitoring radiofrequency (RF) ablation procedures with acoustic radiation force impulse (ARFI) imaging. However, questions remained regarding the utility of the technique in clinically realistic scenarios and at scanning depths associated with abdominal imaging in adults. We address several of these issues and detail recent progress towards the clinical relevance of the ARFI technique. Results from in vitro bovine tissues and an in vivo ovine model are presented. Additional experiments were conducted with a tissue-mimicking phantom and parallel receive tracking techniques in order to further support the clinical feasibility of the method. Thermal lesions created during RF ablation are visualized with high contrast in both in vitro and in vivo hepatic tissues, and radial lesion growth can be monitored throughout the duration of the procedure. ARFI imaging is implemented on a diagnostic ultrasonic scanner, and thus may be a convenient option to guide RF ablation procedures, particularly when electrode insertion is also performed with sonographic guidance.

Duke Scholars

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

August 7, 2006

Volume

51

Issue

15

Start / End Page

3785 / 3808

Location

England

Related Subject Headings

  • Thermal Conductivity
  • Temperature
  • Surgery, Computer-Assisted
  • Sheep
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Liver
  • Image Processing, Computer-Assisted
  • Image Interpretation, Computer-Assisted
  • Hot Temperature
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Fahey, B. J., Hsu, S. J., Wolf, P. D., Nelson, R. C., & Trahey, G. E. (2006). Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities. Phys Med Biol, 51(15), 3785–3808. https://doi.org/10.1088/0031-9155/51/15/013
Fahey, B. J., S. J. Hsu, P. D. Wolf, R. C. Nelson, and G. E. Trahey. “Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities.Phys Med Biol 51, no. 15 (August 7, 2006): 3785–3808. https://doi.org/10.1088/0031-9155/51/15/013.
Fahey BJ, Hsu SJ, Wolf PD, Nelson RC, Trahey GE. Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities. Phys Med Biol. 2006 Aug 7;51(15):3785–808.
Fahey, B. J., et al. “Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities.Phys Med Biol, vol. 51, no. 15, Aug. 2006, pp. 3785–808. Pubmed, doi:10.1088/0031-9155/51/15/013.
Fahey BJ, Hsu SJ, Wolf PD, Nelson RC, Trahey GE. Liver ablation guidance with acoustic radiation force impulse imaging: challenges and opportunities. Phys Med Biol. 2006 Aug 7;51(15):3785–3808.
Journal cover image

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

August 7, 2006

Volume

51

Issue

15

Start / End Page

3785 / 3808

Location

England

Related Subject Headings

  • Thermal Conductivity
  • Temperature
  • Surgery, Computer-Assisted
  • Sheep
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Liver
  • Image Processing, Computer-Assisted
  • Image Interpretation, Computer-Assisted
  • Hot Temperature