Skip to main content
Journal cover image

Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.

Publication ,  Journal Article
Truong, T-K; Chen, N-K; Song, AW
Published in: Neuroimage
August 15, 2011

In diffusion tensor imaging (DTI), spatial and temporal variations of the static magnetic field (B(0)) caused by susceptibility effects and time-varying eddy currents result in severe distortions, blurring, and misregistration artifacts, which in turn lead to errors in DTI metrics and in fiber tractography. Various correction methods have been proposed, but typically assume that the eddy current-induced magnetic field can be modeled as a constant or a single exponential decay within the DTI readout window. Here, we show that its temporal dependence is more complex because of the interaction of multiple eddy currents with different time constants, but that it remains very consistent over time. As such, we propose a novel dynamic B(0) mapping and off-resonance correction method that measures the exact spatial, temporal, and diffusion-weighting direction dependence of the susceptibility- and eddy current-induced magnetic fields to effectively and efficiently correct for artifacts caused by both susceptibility effects and time-varying eddy currents, thereby resulting in a high spatial fidelity and accuracy.

Duke Scholars

Published In

Neuroimage

DOI

EISSN

1095-9572

Publication Date

August 15, 2011

Volume

57

Issue

4

Start / End Page

1343 / 1347

Location

United States

Related Subject Headings

  • Neurology & Neurosurgery
  • Humans
  • Diffusion Tensor Imaging
  • Brain Mapping
  • Artifacts
  • 42 Health sciences
  • 32 Biomedical and clinical sciences
  • 17 Psychology and Cognitive Sciences
  • 11 Medical and Health Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Truong, T.-K., Chen, N.-K., & Song, A. W. (2011). Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging. Neuroimage, 57(4), 1343–1347. https://doi.org/10.1016/j.neuroimage.2011.06.008
Truong, Trong-Kha, Nan-kuei Chen, and Allen W. Song. “Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.Neuroimage 57, no. 4 (August 15, 2011): 1343–47. https://doi.org/10.1016/j.neuroimage.2011.06.008.
Truong, Trong-Kha, et al. “Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.Neuroimage, vol. 57, no. 4, Aug. 2011, pp. 1343–47. Pubmed, doi:10.1016/j.neuroimage.2011.06.008.
Journal cover image

Published In

Neuroimage

DOI

EISSN

1095-9572

Publication Date

August 15, 2011

Volume

57

Issue

4

Start / End Page

1343 / 1347

Location

United States

Related Subject Headings

  • Neurology & Neurosurgery
  • Humans
  • Diffusion Tensor Imaging
  • Brain Mapping
  • Artifacts
  • 42 Health sciences
  • 32 Biomedical and clinical sciences
  • 17 Psychology and Cognitive Sciences
  • 11 Medical and Health Sciences