Single-shot dual-z-shimmed sensitivity-encoded spiral-in/out imaging for functional MRI with reduced susceptibility artifacts.
Blood oxygenation level-dependent (BOLD) functional MRI (fMRI) can be severely hampered by signal loss due to susceptibility-induced static magnetic field (B(0)) inhomogeneities near air/tissue interfaces. A single-shot spiral-in/out sequence with a z-shim gradient embedded between the two acquisitions was previously proposed to efficiently recover the signal. However, despite promising results, this technique had several limitations, which are addressed here as follows. First, by adding a second z-shim gradient before the spiral-in acquisition and optimizing both z-shim gradients slice-by-slice, a significantly more uniform signal recovery can be achieved. Second, by acquiring a B(0) map, the optimal z-shim gradients can be directly, efficiently, and accurately determined for each subject. Third, by complementing the z-shimming approach with sensitivity encoding (SENSE), the in-plane spatial resolution can be increased and, hence, susceptibility artifacts further reduced, while maintaining a high temporal resolution for fMRI applications. These advantages are demonstrated in human functional studies.
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Nuclear Medicine & Medical Imaging
- Magnetic Resonance Imaging
- Image Processing, Computer-Assisted
- Image Enhancement
- Humans
- Brain Mapping
- Artifacts
- 4003 Biomedical engineering
- 0903 Biomedical Engineering
Citation
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- Nuclear Medicine & Medical Imaging
- Magnetic Resonance Imaging
- Image Processing, Computer-Assisted
- Image Enhancement
- Humans
- Brain Mapping
- Artifacts
- 4003 Biomedical engineering
- 0903 Biomedical Engineering