Independent verification of gantry angle for pre-treatment VMAT QA using EPID.

Publication ,  Journal Article
Adamson, J; Wu, Q
Published in: Phys Med Biol
October 21, 2012

We propose a method to incorporate independent verification of gantry angle for electronic portal imaging device (EPID)-based pre-treatment quality assurance (QA) of clinical volumetric modulated arc therapy (VMAT) plans. Gantry angle is measured using projections in the EPID of a custom phantom placed on the couch and the treatment plan is modified so as to be incident on the phantom with a portion of the beam that is collimated in the clinical plan. For our implementation, collimator and couch angles were set to zero and the inferior jaw and two most inferior multi-leaf collimator pairs were opened for the entire QA delivery. A phantom containing five gold coils was used to measure the gantry rotation through which each portal image was acquired. We performed the EPID QA for ten clinical plans and evaluated accuracy of gantry angle measurement, scatter incident on the imager due to the phantom, inter-image pixel linearity and inter- and intra-image noise. The gantry angle could be measured to within 0.0 ± 0.3° for static gantry and 0.2 ± 0.2° for arc acquisitions. Scatter due to the presence of the phantom was negligible. The procedure was shown to be feasible and adds gantry angle to the treatment planning parameters that can be verified by EPID-based pre-treatment VMAT QA.

Duke Scholars

Published In

Phys Med Biol

DOI

EISSN

1361-6560

Publication Date

October 21, 2012

Volume

57

Issue

20

Start / End Page

6587 / 6600

Location

England

Related Subject Headings

  • Radiotherapy, Intensity-Modulated
  • Radiotherapy Planning, Computer-Assisted
  • Quality Control
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Feasibility Studies
  • Electrical Equipment and Supplies
  • 5105 Medical and biological physics
  • 1103 Clinical Sciences
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Adamson, J., & Wu, Q. (2012). Independent verification of gantry angle for pre-treatment VMAT QA using EPID. Phys Med Biol, 57(20), 6587–6600. https://doi.org/10.1088/0031-9155/57/20/6587
Adamson, Justus, and Qiuwen Wu. “Independent verification of gantry angle for pre-treatment VMAT QA using EPID.Phys Med Biol 57, no. 20 (October 21, 2012): 6587–6600. https://doi.org/10.1088/0031-9155/57/20/6587.
Adamson J, Wu Q. Independent verification of gantry angle for pre-treatment VMAT QA using EPID. Phys Med Biol. 2012 Oct 21;57(20):6587–600.
Adamson, Justus, and Qiuwen Wu. “Independent verification of gantry angle for pre-treatment VMAT QA using EPID.Phys Med Biol, vol. 57, no. 20, Oct. 2012, pp. 6587–600. Pubmed, doi:10.1088/0031-9155/57/20/6587.
Adamson J, Wu Q. Independent verification of gantry angle for pre-treatment VMAT QA using EPID. Phys Med Biol. 2012 Oct 21;57(20):6587–6600.
Journal cover image

Published In

Phys Med Biol

DOI

EISSN

1361-6560

Publication Date

October 21, 2012

Volume

57

Issue

20

Start / End Page

6587 / 6600

Location

England

Related Subject Headings

  • Radiotherapy, Intensity-Modulated
  • Radiotherapy Planning, Computer-Assisted
  • Quality Control
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Feasibility Studies
  • Electrical Equipment and Supplies
  • 5105 Medical and biological physics
  • 1103 Clinical Sciences
  • 0903 Biomedical Engineering