Dosimetric characteristics of Novalis shaped beam surgery unit.

Publication ,  Journal Article
Yin, F-F; Zhu, J; Yan, H; Gaun, H; Hammoud, R; Ryu, S; Kim, JH
Published in: Med Phys
August 2002

The dosimetric characteristics of a new dedicated radiosurgical treatment unit are systematically measured in terms of its percent depth dose, beam profile, and relative scatter factor. High-resolution diode detector, mini-ion-chamber detector, and conventional Kodak XV films are used to measure dosimetric data for a range of field sizes from 6x6 mm to 100x100 mm. The effects of collimator size, micro-multileaf collimator shape, and detector type on the dosimetric data are investigated. Results indicate that, with careful design, accurate dosimetric data could be acquired using either a dedicated diode detector or a mini-ion-chamber detector, and film detector. Special attention is required when measuring dosimetric data for small field sizes such as 6x6 mm.

Duke Scholars

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

August 2002

Volume

29

Issue

8

Start / End Page

1729 / 1738

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Semiconductors
  • Scattering, Radiation
  • Reproducibility of Results
  • Radiotherapy, High-Energy
  • Radiotherapy, Conformal
  • Radiotherapy Planning, Computer-Assisted
  • Radiosurgery
  • Radiometry
  • Radiation Dosage
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Yin, F.-F., Zhu, J., Yan, H., Gaun, H., Hammoud, R., Ryu, S., & Kim, J. H. (2002). Dosimetric characteristics of Novalis shaped beam surgery unit. Med Phys, 29(8), 1729–1738. https://doi.org/10.1118/1.1494830
Yin, Fang-Fang, Jingeng Zhu, Hui Yan, Haiqun Gaun, Rabih Hammoud, Samuel Ryu, and Jae H. Kim. “Dosimetric characteristics of Novalis shaped beam surgery unit.Med Phys 29, no. 8 (August 2002): 1729–38. https://doi.org/10.1118/1.1494830.
Yin F-F, Zhu J, Yan H, Gaun H, Hammoud R, Ryu S, et al. Dosimetric characteristics of Novalis shaped beam surgery unit. Med Phys. 2002 Aug;29(8):1729–38.
Yin, Fang-Fang, et al. “Dosimetric characteristics of Novalis shaped beam surgery unit.Med Phys, vol. 29, no. 8, Aug. 2002, pp. 1729–38. Pubmed, doi:10.1118/1.1494830.
Yin F-F, Zhu J, Yan H, Gaun H, Hammoud R, Ryu S, Kim JH. Dosimetric characteristics of Novalis shaped beam surgery unit. Med Phys. 2002 Aug;29(8):1729–1738.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

August 2002

Volume

29

Issue

8

Start / End Page

1729 / 1738

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Semiconductors
  • Scattering, Radiation
  • Reproducibility of Results
  • Radiotherapy, High-Energy
  • Radiotherapy, Conformal
  • Radiotherapy Planning, Computer-Assisted
  • Radiosurgery
  • Radiometry
  • Radiation Dosage