Improved spatial resolution for cardiac mapping using current source density-based electrode arrays.

Publication ,  Journal Article
Pieper, CF; Pacifico, A
Published in: Pacing Clin Electrophysiol
January 1995

The time of the minimum slope (i.e., the fastest negative deflection) in monopolar (MP) electrograms from normal hearts compares closely with time of phase 0 of the action potential in cells underlying the electrode, but poor rejection of far-field activity may limit the utility of MP electrode technology in dense arrays used for the study of ventricular tachycardia and fibrillation. The purpose of this study is to evaluate more myopic discrete bipolar (BP) and nondirectional, two-dimensional current source density (CSD) based arrays for rejection of far-field potentials and precision of activation time determination. Simultaneous recordings of the CSD, MP, and multiple BP electrograms were performed on normal dog epicardium. The time of the minimum slope in MP electrograms was compared to activation times in CSD and BP derivations using: (1) peak; (2) steepest slope; (3) zero crossing of the steepest sloping segment in either direction; and (4) waveform morphology. In vivo, CSD amplitude was reduced significantly more than MP and BP amplitudes by insertion of inert media between the heart and the electrodes. The time of the steepest slope in CSD electrograms designated activation times closest to the time of the minimum slope in MP electrograms (0.9 +/- 1.3 msec). We conclude that CSD provides a nondirectional electrode system that accurately defines the time of local activation and possesses better spatial specificity than MP electrode systems and BP electrode systems having the same interelectrode distances.

Duke Scholars

Published In

Pacing Clin Electrophysiol

DOI

ISSN

0147-8389

Publication Date

January 1995

Volume

18

Issue

1 Pt 1

Start / End Page

34 / 44

Location

United States

Related Subject Headings

  • Male
  • Heart Conduction System
  • Female
  • Electrodes, Implanted
  • Electrocardiography
  • Dogs
  • Cardiovascular System & Hematology
  • Animals
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Pieper, C. F., & Pacifico, A. (1995). Improved spatial resolution for cardiac mapping using current source density-based electrode arrays. Pacing Clin Electrophysiol, 18(1 Pt 1), 34–44. https://doi.org/10.1111/j.1540-8159.1995.tb02473.x
Pieper, C. F., and A. Pacifico. “Improved spatial resolution for cardiac mapping using current source density-based electrode arrays.Pacing Clin Electrophysiol 18, no. 1 Pt 1 (January 1995): 34–44. https://doi.org/10.1111/j.1540-8159.1995.tb02473.x.
Pieper CF, Pacifico A. Improved spatial resolution for cardiac mapping using current source density-based electrode arrays. Pacing Clin Electrophysiol. 1995 Jan;18(1 Pt 1):34–44.
Pieper, C. F., and A. Pacifico. “Improved spatial resolution for cardiac mapping using current source density-based electrode arrays.Pacing Clin Electrophysiol, vol. 18, no. 1 Pt 1, Jan. 1995, pp. 34–44. Pubmed, doi:10.1111/j.1540-8159.1995.tb02473.x.
Pieper CF, Pacifico A. Improved spatial resolution for cardiac mapping using current source density-based electrode arrays. Pacing Clin Electrophysiol. 1995 Jan;18(1 Pt 1):34–44.
Journal cover image

Published In

Pacing Clin Electrophysiol

DOI

ISSN

0147-8389

Publication Date

January 1995

Volume

18

Issue

1 Pt 1

Start / End Page

34 / 44

Location

United States

Related Subject Headings

  • Male
  • Heart Conduction System
  • Female
  • Electrodes, Implanted
  • Electrocardiography
  • Dogs
  • Cardiovascular System & Hematology
  • Animals
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology