Anthropomorphic cardiac ultrasound phantom.
A new phantom is described which simulates the human cardiac anatomy for applications in ultrasound imaging, ultrasound Doppler, and color-flow Doppler imaging. The phantom consists of a polymer left ventricle which includes a prosthetic mitral and aortic valve and is connected to a mock circulatory loop. Aerated tap water serves as a blood simulating fluid and ultrasound contrast medium within the circulatory loop. The left ventricle is housed in a Lexan ultrasound visualization chamber which includes ultrasound viewing ports and acoustic absorbers. A piston pump connected to the visualization chamber by a single port pumps degassed water within the chamber which in turn pumps the left ventricle. Real-time ultrasound images and Doppler studies measure flow patterns through the valves and within the left ventricle.
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
DOI
EISSN
ISSN
Publication Date
Volume
Issue
Start / End Page
Related Subject Headings
- Ultrasonography
- Models, Structural
- Humans
- Heart Ventricles
- Heart
- Biomedical Engineering
- 4603 Computer vision and multimedia computation
- 4009 Electronics, sensors and digital hardware
- 4003 Biomedical engineering
- 0906 Electrical and Electronic Engineering
Citation
Published In
DOI
EISSN
ISSN
Publication Date
Volume
Issue
Start / End Page
Related Subject Headings
- Ultrasonography
- Models, Structural
- Humans
- Heart Ventricles
- Heart
- Biomedical Engineering
- 4603 Computer vision and multimedia computation
- 4009 Electronics, sensors and digital hardware
- 4003 Biomedical engineering
- 0906 Electrical and Electronic Engineering