Skip to main content

Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy.

Publication ,  Journal Article
Glass, PS; Jacobs, JR; Smith, LR; Ginsberg, B; Quill, TJ; Bai, SA; Reves, JG
Published in: Anesthesiology
December 1990

Computer-assisted continuous infusion (CACI) is a pharmacokinetic model-driven infusion device that enables physicians to administer intravenous (iv) drugs in a quantitative fashion, specifying a theoretical blood or plasma concentration. This study evaluated the accuracy of CACI administration of fentanyl using a newly developed CACI device programmed with a well-known set of pharmacokinetic parameters for fentanyl. Patients received diazepam 1 or 2 h before surgery. Anesthesia was induced by a combination of 70% N2O and fentanyl administered by CACI to a predicted concentration of 15-25 ng.ml-1. After neuromuscular blockade and tracheal intubation, the desired plasma fentanyl concentration (setpoint) entered into CACI was 3-6 ng.ml-1, and then the setpoint fentanyl concentration was titrated according to strict criteria of adequate or inadequate anesthesia. Plasma samples for subsequent assay of fentanyl concentration then were taken: at predefined stimuli, when inadequate anesthesia occurred, or 5 min before an anticipated decrease in the fentanyl setpoint. The predictive accuracy of CACI was assessed by calculating for each patient the tenth, 50th, and 90th percentile of the performance error and absolute performance error from each measured and predicted plasma sample pair. Cumulative probability functions for each of these were then plotted. Precision was defined as the dispersion of the tenth to 90th percentile of the median percent performance error for the population and was found to be -31-26%. The median population performance error was -4%, and the median population absolute performance error was 21%.(ABSTRACT TRUNCATED AT 250 WORDS)

Duke Scholars

Published In

Anesthesiology

DOI

ISSN

0003-3022

Publication Date

December 1990

Volume

73

Issue

6

Start / End Page

1082 / 1090

Location

United States

Related Subject Headings

  • Software
  • Middle Aged
  • Microcomputers
  • Infusions, Intravenous
  • Infusion Pumps
  • Humans
  • Hemodynamics
  • Fentanyl
  • Evaluation Studies as Topic
  • Anesthesiology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Glass, P. S., Jacobs, J. R., Smith, L. R., Ginsberg, B., Quill, T. J., Bai, S. A., & Reves, J. G. (1990). Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy. Anesthesiology, 73(6), 1082–1090. https://doi.org/10.1097/00000542-199012000-00004
Glass, P. S., J. R. Jacobs, L. R. Smith, B. Ginsberg, T. J. Quill, S. A. Bai, and J. G. Reves. “Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy.Anesthesiology 73, no. 6 (December 1990): 1082–90. https://doi.org/10.1097/00000542-199012000-00004.
Glass PS, Jacobs JR, Smith LR, Ginsberg B, Quill TJ, Bai SA, et al. Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy. Anesthesiology. 1990 Dec;73(6):1082–90.
Glass, P. S., et al. “Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy.Anesthesiology, vol. 73, no. 6, Dec. 1990, pp. 1082–90. Pubmed, doi:10.1097/00000542-199012000-00004.
Glass PS, Jacobs JR, Smith LR, Ginsberg B, Quill TJ, Bai SA, Reves JG. Pharmacokinetic model-driven infusion of fentanyl: assessment of accuracy. Anesthesiology. 1990 Dec;73(6):1082–1090.

Published In

Anesthesiology

DOI

ISSN

0003-3022

Publication Date

December 1990

Volume

73

Issue

6

Start / End Page

1082 / 1090

Location

United States

Related Subject Headings

  • Software
  • Middle Aged
  • Microcomputers
  • Infusions, Intravenous
  • Infusion Pumps
  • Humans
  • Hemodynamics
  • Fentanyl
  • Evaluation Studies as Topic
  • Anesthesiology