Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.
Publication
, Journal Article
Sheth, N; Mahmood, K; Singh, B; Carter-Adkins, D; Pachulski, RT
Published in: Can J Cardiol
April 2002
The lead conductor integrity of implantable cardioverter defibrillator devices is inferred from impedance measurements; however, intermittent discontinuity can be difficult to detect or confirm. Newer devices can perform daily lead impedance self-testing, and some even have audible alarms that promptly warn patients of anomalies. In the present case, the audible alarms were solely responsible for the timely identification of an intermittent, otherwise clinically nonreproducible, form of potentially fatal implantable cardioverter defibrillator system failure.
Duke Scholars
Published In
Can J Cardiol
ISSN
0828-282X
Publication Date
April 2002
Volume
18
Issue
4
Start / End Page
430 / 432
Location
England
Related Subject Headings
- Humans
- Female
- Equipment Failure
- Defibrillators, Implantable
- Death, Sudden, Cardiac
- Cardiovascular System & Hematology
- Cardiomyopathies
- Cardiac Pacing, Artificial
- Arrhythmias, Cardiac
- Aged
Citation
APA
Chicago
ICMJE
MLA
NLM
Sheth, N., Mahmood, K., Singh, B., Carter-Adkins, D., & Pachulski, R. T. (2002). Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol, 18(4), 430–432.
Sheth, Nirav, Kamran Mahmood, Binod Singh, Danielle Carter-Adkins, and Roman T. Pachulski. “Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.” Can J Cardiol 18, no. 4 (April 2002): 430–32.
Sheth N, Mahmood K, Singh B, Carter-Adkins D, Pachulski RT. Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol. 2002 Apr;18(4):430–2.
Sheth, Nirav, et al. “Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.” Can J Cardiol, vol. 18, no. 4, Apr. 2002, pp. 430–32.
Sheth N, Mahmood K, Singh B, Carter-Adkins D, Pachulski RT. Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol. 2002 Apr;18(4):430–432.
Published In
Can J Cardiol
ISSN
0828-282X
Publication Date
April 2002
Volume
18
Issue
4
Start / End Page
430 / 432
Location
England
Related Subject Headings
- Humans
- Female
- Equipment Failure
- Defibrillators, Implantable
- Death, Sudden, Cardiac
- Cardiovascular System & Hematology
- Cardiomyopathies
- Cardiac Pacing, Artificial
- Arrhythmias, Cardiac
- Aged