Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.

Publication ,  Journal Article
Sheth, N; Mahmood, K; Singh, B; Carter-Adkins, D; Pachulski, RT
Published in: Can J Cardiol
April 2002

The lead conductor integrity of implantable cardioverter defibrillator devices is inferred from impedance measurements; however, intermittent discontinuity can be difficult to detect or confirm. Newer devices can perform daily lead impedance self-testing, and some even have audible alarms that promptly warn patients of anomalies. In the present case, the audible alarms were solely responsible for the timely identification of an intermittent, otherwise clinically nonreproducible, form of potentially fatal implantable cardioverter defibrillator system failure.

Duke Scholars

Published In

Can J Cardiol

ISSN

0828-282X

Publication Date

April 2002

Volume

18

Issue

4

Start / End Page

430 / 432

Location

England

Related Subject Headings

  • Humans
  • Female
  • Equipment Failure
  • Defibrillators, Implantable
  • Death, Sudden, Cardiac
  • Cardiovascular System & Hematology
  • Cardiomyopathies
  • Cardiac Pacing, Artificial
  • Arrhythmias, Cardiac
  • Aged
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Sheth, N., Mahmood, K., Singh, B., Carter-Adkins, D., & Pachulski, R. T. (2002). Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol, 18(4), 430–432.
Sheth, Nirav, Kamran Mahmood, Binod Singh, Danielle Carter-Adkins, and Roman T. Pachulski. “Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.Can J Cardiol 18, no. 4 (April 2002): 430–32.
Sheth N, Mahmood K, Singh B, Carter-Adkins D, Pachulski RT. Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol. 2002 Apr;18(4):430–2.
Sheth, Nirav, et al. “Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity.Can J Cardiol, vol. 18, no. 4, Apr. 2002, pp. 430–32.
Sheth N, Mahmood K, Singh B, Carter-Adkins D, Pachulski RT. Audible implantable cardioverter defibrillator alarms detect intermittent conductor discontinuity. Can J Cardiol. 2002 Apr;18(4):430–432.
Journal cover image

Published In

Can J Cardiol

ISSN

0828-282X

Publication Date

April 2002

Volume

18

Issue

4

Start / End Page

430 / 432

Location

England

Related Subject Headings

  • Humans
  • Female
  • Equipment Failure
  • Defibrillators, Implantable
  • Death, Sudden, Cardiac
  • Cardiovascular System & Hematology
  • Cardiomyopathies
  • Cardiac Pacing, Artificial
  • Arrhythmias, Cardiac
  • Aged