Evaluation of an electron Monte Carlo dose calculation algorithm for electron beams
The electron Monte Carlo (eMC) dose calculation algorithm of the Eclipse treatment planning system is based heavily upon Monte Carlo simulation of the linac head and modeling of the linac beam characteristics with minimal measurement of beam data. Commissioning of the eMC algorithm on multiple identical linacs provided a unique opportunity to systematically evaluate the algorithm with actual measurements of clinically relevant beam and dose parameters. In this study, measured and eMC calculated dose distributions were compared both along and perpendicular to electron beam direction for electron energy/applicator/ depth combination using measurement data from four Varian CLINAC 21EX linear accelerators (Varian Medical Systems, Palo Alto, CA). Cutout factors for sizes down to 3 × 3 cm were also compared. Comparisons between the measurement and the eMC calculated values show that the R