SU‐FF‐I‐17: Reduction of Set‐Up Error Using Cone Beam‐CT in Patients Undergoing Partial Breast Irradiation

Publication ,  Conference
Fatunase, T; Wang, Z; Yoo, S; Yin, F; Marks, L
Published in: Medical Physics
January 1, 2006

Purpose: Partial breast irradiation with multiple external beams requires accurate alignment of the target volume and treatment isocenter. On‐board plannar kV/MV imaging can verify set‐up based on boney landmarks. On‐board cone beam CT (CBCT) may provide additional soft‐tissue‐based information to further improve set‐up accuracy. In this study, we assess the utility of CBCT in patients receiving partial breast radiation therapy, who have been aligned with on‐board plannar kV/MV imaging. Method and Materials: Patients undergoing partial breast irradiation were imaged using an on‐board‐imager attached to the gantry of a Varian 21EX machine. Each patient was aligned to skin marks in the treatment position. Orthogonal kV images were acquired and registered to digitally reconstructed radiographs from the planning CT. Subsequently, a CBCT image data set was acquired and compared to the planning CT using both boney anatomy and soft tissue information which yielded estimations of residual setup error. Results: 10 patients were studied under an IRB protocol. Each patient had an average acquisition of 9 pre‐RT CBCT images. After 2D kV image registration based on boney anatomy, the average soft tissue residual error, based on the CBCT vs. planning CT, was 4, 2, and 3 mm in the Ant/Post, Sup/Inf, and Rt/Lt directions respectively. The ranges were 0–15, 0–9, and 0–7 mm, respectively. In one case, 30% of the planning tumor volume (PTV) would have been outside of the therapeutic isodose volume based on the pre‐RT CBCT images. Conclusion: CBCT is a practical tool for image registration while visualizing the soft tissue of the tumor bed. It provides additional anatomic information and soft‐tissue detail beyond that provided by planar radiographs. Such increased accuracy can significantly improve dosimetric coverage of the planning target volume. Conflict of Interest: Partially supported by a Varian research grant. © 2006, American Association of Physicists in Medicine. All rights reserved.

Duke Scholars

Published In

Medical Physics

DOI

ISSN

0094-2405

Publication Date

January 1, 2006

Volume

33

Issue

6

Start / End Page

2000

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Fatunase, T., Wang, Z., Yoo, S., Yin, F., & Marks, L. (2006). SU‐FF‐I‐17: Reduction of Set‐Up Error Using Cone Beam‐CT in Patients Undergoing Partial Breast Irradiation. In Medical Physics (Vol. 33, p. 2000). https://doi.org/10.1118/1.2240255
Fatunase, T., Z. Wang, S. Yoo, F. Yin, and L. Marks. “SU‐FF‐I‐17: Reduction of Set‐Up Error Using Cone Beam‐CT in Patients Undergoing Partial Breast Irradiation.” In Medical Physics, 33:2000, 2006. https://doi.org/10.1118/1.2240255.
Fatunase T, Wang Z, Yoo S, Yin F, Marks L. SU‐FF‐I‐17: Reduction of Set‐Up Error Using Cone Beam‐CT in Patients Undergoing Partial Breast Irradiation. In: Medical Physics. 2006. p. 2000.
Fatunase, T., et al. “SU‐FF‐I‐17: Reduction of Set‐Up Error Using Cone Beam‐CT in Patients Undergoing Partial Breast Irradiation.” Medical Physics, vol. 33, no. 6, 2006, p. 2000. Scopus, doi:10.1118/1.2240255.

Published In

Medical Physics

DOI

ISSN

0094-2405

Publication Date

January 1, 2006

Volume

33

Issue

6

Start / End Page

2000

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences