Skip to main content

Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters

Publication ,  Journal Article
Giaddui, T; Cui, Y; Galvin, J; Yu, Y; Xiao, Y
Published in: Ifmbe Proceedings
April 16, 2013

Surface doses received during seven different imaging protocols (using the kV XVI imager) were measured on a Rando phantom surface using nanoDot optical stimulated luminescence dosimeters (OSLD) for three different body regions (head and neck, chest and pelvis). For each protocol, the surface dose was measured at four different locations on the surface of the phantom (ANT., POST., LLAT. and RLAT.). The surface dose at any location in the irradiated area can range between 0.008 cGy (fast head and neck protocol) and 4.38 cGy (symmetry 4D). The average surface dose in the irradiated area ranged between 0.038 cGy and 2.34 cGy. The measured doses were compared with nominal scan dose, provided by the vendor and calculated doses. © 2013 Springer-Verlag.

Duke Scholars

Published In

Ifmbe Proceedings

DOI

ISSN

1680-0737

Publication Date

April 16, 2013

Volume

39 IFMBE

Start / End Page

1195 / 1198
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Giaddui, T., Cui, Y., Galvin, J., Yu, Y., & Xiao, Y. (2013). Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters. Ifmbe Proceedings, 39 IFMBE, 1195–1198. https://doi.org/10.1007/978-3-642-29305-4_313
Giaddui, T., Y. Cui, J. Galvin, Y. Yu, and Y. Xiao. “Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters.” Ifmbe Proceedings 39 IFMBE (April 16, 2013): 1195–98. https://doi.org/10.1007/978-3-642-29305-4_313.
Giaddui T, Cui Y, Galvin J, Yu Y, Xiao Y. Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters. Ifmbe Proceedings. 2013 Apr 16;39 IFMBE:1195–8.
Giaddui, T., et al. “Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters.” Ifmbe Proceedings, vol. 39 IFMBE, Apr. 2013, pp. 1195–98. Scopus, doi:10.1007/978-3-642-29305-4_313.
Giaddui T, Cui Y, Galvin J, Yu Y, Xiao Y. Surface dose measurements of kV XVI cone-beam CT system using NanoDot optical stimulated luminescence dosimeters. Ifmbe Proceedings. 2013 Apr 16;39 IFMBE:1195–1198.

Published In

Ifmbe Proceedings

DOI

ISSN

1680-0737

Publication Date

April 16, 2013

Volume

39 IFMBE

Start / End Page

1195 / 1198