Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance.

Publication ,  Journal Article
Ghate, SV; Baker, JA; Hawkins, AD; Soher, BJ
Published in: Acad Radiol
June 2011

RATIONALE AND OBJECTIVES: Spectroscopy signal of a breast lesion may be disrupted by the presence of metal from a biopsy marking clip. This study compares the size of magnetic resonance (MR) susceptibility artifacts and degree of local spectroscopy signal disturbance created by conventional titanium tissue marker clips to that of a novel carbon coated ceramic breast tissue marker clip. MATERIALS AND METHODS: Five breast tissue marker clips were embedded in a gelatin breast phantom. The phantoms were imaged on a 3T MR scanner, and three-dimensional T1-weighted gradient echo images were obtained. The area of the susceptibility artifact was calculated and compared for each clip. Single voxel point resolved spectroscopy spectra (SVS) were acquired for three ceramic clips aligned along the superoinferior, anteroposterior, and left-right axes, respectively. Measurements were repeated for an area of pure gelatin and for one similarly sized titanium clip located arbitrarily in the main field. Water spectra were obtained, and line widths and areas of the water peaks were compared. RESULTS: All five clips were easily visible on MR imaging. The ceramic marker clip produced less apparent artifact when compared with the metallic clips. Spectral analysis demonstrated local frequency shifts around all clips. Line widths varied from 7 to 92 Hz in the voxels on and around the titanium clip and from 4 to 18 Hz around the ceramic clips. CONCLUSION: The ceramic breast tissue marker clip produced less susceptibility artifact and less line broadening on 3T MR imaging than conventional titanium clips. This tissue marker may reduce artifact and improve consistency of breast MR spectroscopy.

Duke Scholars

Published In

Acad Radiol

DOI

EISSN

1878-4046

Publication Date

June 2011

Volume

18

Issue

6

Start / End Page

770 / 773

Location

United States

Related Subject Headings

  • Titanium
  • Prostheses and Implants
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Female
  • Ceramics
  • Carbon
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Ghate, S. V., Baker, J. A., Hawkins, A. D., & Soher, B. J. (2011). Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance. Acad Radiol, 18(6), 770–773. https://doi.org/10.1016/j.acra.2011.01.008
Ghate, Sujata V., Jay A. Baker, Ashley D. Hawkins, and Brian J. Soher. “Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance.Acad Radiol 18, no. 6 (June 2011): 770–73. https://doi.org/10.1016/j.acra.2011.01.008.
Ghate, Sujata V., et al. “Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance.Acad Radiol, vol. 18, no. 6, June 2011, pp. 770–73. Pubmed, doi:10.1016/j.acra.2011.01.008.
Ghate SV, Baker JA, Hawkins AD, Soher BJ. Titanium vs carbon coated ceramic breast tissue marker clips: 3T MR susceptibility artifact and local signal disturbance. Acad Radiol. 2011 Jun;18(6):770–773.
Journal cover image

Published In

Acad Radiol

DOI

EISSN

1878-4046

Publication Date

June 2011

Volume

18

Issue

6

Start / End Page

770 / 773

Location

United States

Related Subject Headings

  • Titanium
  • Prostheses and Implants
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Female
  • Ceramics
  • Carbon