Skip to main content
Journal cover image

A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery.

Publication ,  Journal Article
Grewal, DS; Stinnett, SS; Folgar, FA; Schneider, EW; Vajzovic, L; Asrani, S; Freedman, SF; Mruthyunjaya, P; Hahn, P
Published in: Am J Ophthalmol
January 2016

PURPOSE: To investigate agreement in intraocular pressure (IOP) measurements among Icare rebound tonometry, Tonopen tonometry, and Goldmann applanation tonometry following vitreoretinal surgery. DESIGN: Reliability analysis of tonometers. METHODS: Fifty-eyes of 50 adults undergoing vitreoretinal surgery were enrolled. IOP was measured on first postoperative day using Icare (Tiolat, Helsinki, Finland), followed by Tonopen (Reichert, Depew, New York, USA) and Goldmann (Haag-Streit USA, Mason, Ohio, USA) in randomized order. Intraclass correlation coefficients (ICC) and Bland-Altman plots were calculated for all subcategories. RESULTS: Icare successfully measured IOP in all eyes, while Goldmann was unmeasurable in 6 eyes (12%) and Tonopen in 1 eye (2%). Mean IOP by Icare, Tonopen, and Goldmann was 15.9 ± 8.9, 16.9 ± 6.2, and 16.0 ± 7.3 mm Hg, respectively (P = .76). Type of intraocular tamponade, status of lens, status of cornea, gauge of instrumentation, and history of prior vitrectomy did not result in significant differences among the 3 tonometers. ICC was excellent (>0.75) in all subgroups, except at IOP <10 and ≥ 23 mm Hg (based on Icare). In eyes with IOP <10 mm Hg, Icare underestimated IOP (mm Hg; P = .01) compared to Goldmann (2.0 ± 2.1) and Tonopen (3.5 ± 2.4), whereas at IOP ≥ 23 mm Hg Icare was overestimated (P = .01) compared to Goldmann (3.77 ± 3.49) and Tonopen (4.97 ± 3.33). Overall, differences in IOP were ≤ 3 mm Hg in 58% of eyes for Icare-Tonopen, 72% for Tonopen-Goldmann, and 62% for Icare-Goldmann. CONCLUSION: IOP measurements using Icare rebound tonometry, Tonopen, and Goldmann tonometry are in excellent agreement following vitreoretinal surgery. However, Icare overestimates at IOP ≥ 23 and underestimates at IOP <10 mm Hg.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Am J Ophthalmol

DOI

EISSN

1879-1891

Publication Date

January 2016

Volume

161

Start / End Page

22-8.e1-8

Location

United States

Related Subject Headings

  • Vitreoretinal Surgery
  • Tonometry, Ocular
  • Reproducibility of Results
  • Ophthalmology & Optometry
  • Middle Aged
  • Male
  • Intraocular Pressure
  • Humans
  • Female
  • Aged
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Grewal, D. S., Stinnett, S. S., Folgar, F. A., Schneider, E. W., Vajzovic, L., Asrani, S., … Hahn, P. (2016). A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery. Am J Ophthalmol, 161, 22-8.e1-8. https://doi.org/10.1016/j.ajo.2015.09.022
Grewal, Dilraj S., Sandra S. Stinnett, Francisco A. Folgar, Eric W. Schneider, Lejla Vajzovic, Sanjay Asrani, Sharon F. Freedman, Prithvi Mruthyunjaya, and Paul Hahn. “A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery.Am J Ophthalmol 161 (January 2016): 22-8.e1-8. https://doi.org/10.1016/j.ajo.2015.09.022.
Grewal DS, Stinnett SS, Folgar FA, Schneider EW, Vajzovic L, Asrani S, et al. A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery. Am J Ophthalmol. 2016 Jan;161:22-8.e1-8.
Grewal, Dilraj S., et al. “A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery.Am J Ophthalmol, vol. 161, Jan. 2016, pp. 22-8.e1-8. Pubmed, doi:10.1016/j.ajo.2015.09.022.
Grewal DS, Stinnett SS, Folgar FA, Schneider EW, Vajzovic L, Asrani S, Freedman SF, Mruthyunjaya P, Hahn P. A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery. Am J Ophthalmol. 2016 Jan;161:22-8.e1-8.
Journal cover image

Published In

Am J Ophthalmol

DOI

EISSN

1879-1891

Publication Date

January 2016

Volume

161

Start / End Page

22-8.e1-8

Location

United States

Related Subject Headings

  • Vitreoretinal Surgery
  • Tonometry, Ocular
  • Reproducibility of Results
  • Ophthalmology & Optometry
  • Middle Aged
  • Male
  • Intraocular Pressure
  • Humans
  • Female
  • Aged